Settings
TruLab pH 1310
42
ba76156e01
03/2014
*
Brand names or trade names are trademarks of their respective owners pro-
tected by law.
7.1.3 Calibration interval
The calibration evaluation is displayed as a sensor symbol.
The sensor symbol flashes after the adjusted calibration interval has expired. It
is still possible to measure.
Setting the calibra-
tion interval
The calibration interval is set to 7 days in the factory.
You can change the interval (1 ... 999 days):
7.2
Settings for ORP measurements
The settings are made in the menu for calibration and measurement settings of
the pH/ORP measurement. To open the settings, display the required parame-
23
Reagecon 25
*
2.000
4.000
7.000
10.000
13.000
25 °C
24
Chemsolute
*
2,000
4,000
7,000
10,000
20 °C
25
USABlueBook
*
4,000
7,000
10,000
25 °C
No.
Buffer set *
pH values
at
The buffers are selected in the menu, pH /
<F1>
/[Menu] /
Calibra-
tion
/
Buffer
(see 7.1.1 C
To ensure the high measuring accuracy of the measuring system,
calibrate after the calibration interval has expired.
1.
Open the menu for measurement settings with [Menu]
<F1>
.
2.
In the
Calibration
/
Calibration interval
menu, set the calibration interval
with
<
><
>
.
3.
Confirm the setting with
<ENTER>
.
4.
Quit the menu with
<M>
.