pH value
TruLab pH/ISE 1320
38
ba76161e01
08/2015
6.2.6 Calibration data
Displaying the
calibration data
The calibration record of the last calibration is to be found under the menu item,
<MENU/ENTER>
/
Calibration
/ /
Calibration record
. To open it, press the
<CAL__>
key in the measured value display.
Subsequently, you can transmit the displayed calibration data to the interface,
e. g. to a PC, with
<F2>
[USB output]
.
Displaying the
calibration data
storage
The calibration records of the last calibrations (up to 10) are given in the menu,
<MENU/ENTER>
/
Calibration
/
Calibration data storage
.
Calibration
evaluation
After calibrating, the meter automatically evaluates the calibration. The zero
point and slope are evaluated separately. The worse evaluation of both is taken
into account. The evaluation appears on the display and in the calibration
record.
The calibration record is automatically transmitted to the interface
after calibrating.
Menu item
Setting/func-
tion
Description
Calibration
/
Calibration data stor-
age
/
Display
-
Displays the calibration record.
Further options:
Scroll through the calibration
records with
<
><
>
.
Output the displayed calibra-
tion record to the interface with
<F2>
/
[USB output]
.
Quit the display with
<F1>
/
[Back]
or
<MENU/ENTER>
.
Switch directly to the mea-
sured value display with
<M>
.
Calibration
/
Calibration data stor-
age
/
Output to USB
-
Outputs the calibration records to
the interface.