background image

Pro

10 

Quick Start Guide

Item #605193

This Quick Start Guide is meant to serve as a quick reference in calibrating and 
operating the Pro10. It is not intended to replace the information found in the User 
Manual.

Getting Started

Unpack the instrument and install (2) C size batteries in the back of the instrument. 
Tighten the four screws of the battery cover on to the back of the instrument.

Installing the Sensor

Prior to use, install either a pH or ORP sensor into the 60510 cable assembly. Remove 
the temporary plastic plug from the cable’s port by pulling it straight out of the port. 
Remove the plastic plug from the sensor’s connector by pulling it straight off the sensor. 
Insert the sensor into the port so the o-rings on the sensor are no longer visible. Hand 
tightening the sensor into place by twisting it clockwise. Do not use a tool and do not 
over tighten

Enable the correct ISE (either pH or ORP) in the instrument’s System Setup menu. Attach 
the cable assembly to your instrument.

pH Calibration

The Pro10 pH sensor can be calibrated by performing a 1, 2 or 3-point calibration. At 
least one of the calibration points must be done with pH buffer 7 or 6.86. For highest 
accuracy, perform at least a 2-point calibration.

2 or 3-Point Calibration

1. 

Place the sensor in pH buffer 7 or 6.86 and allow the temperature and pH 
readings to stabilize. 

2. 

Press and hold 

Cal

  for three seconds. 

3. 

Highlight 

pH

 and press 

Enter

. If pH is not listed as an option, check the System 

Setup menu to ensure pH is enabled in the ISE Sensor Type menu.

4. 

Highlight 

2 point

 or 

3 point

 and press enter.

5. 

If necessary, use the up and down arrow keys to adjust the pH buffer value.

6. 

Press 

Enter

 to continue to second point.