Pro
10
Quick Start Guide
Item #605193
This Quick Start Guide is meant to serve as a quick reference in calibrating and
operating the Pro10. It is not intended to replace the information found in the User
Manual.
Getting Started
Unpack the instrument and install (2) C size batteries in the back of the instrument.
Tighten the four screws of the battery cover on to the back of the instrument.
Installing the Sensor
Prior to use, install either a pH or ORP sensor into the 60510 cable assembly. Remove
the temporary plastic plug from the cable’s port by pulling it straight out of the port.
Remove the plastic plug from the sensor’s connector by pulling it straight off the sensor.
Insert the sensor into the port so the o-rings on the sensor are no longer visible. Hand
tightening the sensor into place by twisting it clockwise. Do not use a tool and do not
over tighten
Enable the correct ISE (either pH or ORP) in the instrument’s System Setup menu. Attach
the cable assembly to your instrument.
pH Calibration
The Pro10 pH sensor can be calibrated by performing a 1, 2 or 3-point calibration. At
least one of the calibration points must be done with pH buffer 7 or 6.86. For highest
accuracy, perform at least a 2-point calibration.
2 or 3-Point Calibration
1.
Place the sensor in pH buffer 7 or 6.86 and allow the temperature and pH
readings to stabilize.
2.
Press and hold
Cal
for three seconds.
3.
Highlight
pH
and press
Enter
. If pH is not listed as an option, check the System
Setup menu to ensure pH is enabled in the ISE Sensor Type menu.
4.
Highlight
2 point
or
3 point
and press enter.
5.
If necessary, use the up and down arrow keys to adjust the pH buffer value.
6.
Press
Enter
to continue to second point.