Pro

1020 

Quick Start Guide

Item #605188

This Quick Start Guide is meant to serve as a quick reference in calibrating and 
operating the Pro1020. It is not intended to replace the information found in the User 
Manual.

Getting Started

Unpack the instrument and install (2) C size batteries in the back of the instrument. 
Tighten the four screws of the battery plate on to the back of the instrument.

Installing the DO Sensor

Prior to use, install the sensors into the cable assembly. Remove the temporary plastic 
plugs from the cable’s ports by pulling them straight out of the port. Remove the plastic 
plug from the sensor’s connector by pulling it straight off the sensor. Install the sensor 
by inserting the sensor into the port and then hand tightening it. Do not use a tool and 
do not over tighten. Install a pH, pH-amplified or ORP sensor in port 1 and a DO sensor 
in port 2. Install a port plug into any port that does not have an installed sensor. 

Enable the correct ISE (either pH or ORP) and Dissolved Oxygen (either galvanic - gray 
sensor body or polarographic - black sensor body) sensor in the instrument’s System 
Setup menu. Attach the cable assembly to your instrument.

Installing the DO Membrane

The DO sensor is shipped with a red protective cap to protect the electrode.  A new 
membrane cap must be installed before the first use.

1. 

Prepare the O2 probe solution according to the instructions on the bottle. After 
mixing, allow the solution to sit for 1 hour. This will help prevent air bubbles 
from later developing under the membrane. 

2. 

Remove, and discard or save the red protective cap. 

3. 

Thoroughly rinse the sensor tip with distilled or deionized water.

4. 

Fill a new membrane cap with probe solution. Avoid touching the membrane 
portion of the cap.  

5. 

Thread the membrane cap onto the sensor, moderately tight. A small amount of 
electrolyte will overflow.