Taking Measurements

1. 

Before taking measurements, be sure the instrument has been calibrated to 
ensure the most accurate readings.  

2. 

Turn the instrument on and wait 5-15 minutes if using a polarographic sensor.

3. 

Install the sensor guard and then insert the probe into the sample.

4. 

Ensure the conductivity sensor is completely submerged in the sample. The two 
holes near the cable should be covered by the sample for accurate conductivity 
readings.  

5. 

Move the probe in the sample at a rate of at least 6 inches (16 cm) per second if 
using the yellow membrane and 3 inches (7.62 cm) per second if using the blue 
membrane.

6. 

While continuing to provide sample movement, wait for the readings to 
stabilize. 

7. 

Highlight 

Save

 and press 

Enter

 to store the reading. The instrument will confirm 

that the reading was successfully saved.

Contact Information

6. 

If Quick DO Cal is not enabled, highlight 

%

 or 

%Local

 and press 

Enter

. The 

Pro1020 will display the current DO% and temperature readings along with 
the % calibration value. Wait at least 3 seconds, then, once the DO% and 
temperature readings are stable, press 

Enter

 to complete the calibration. 

7. 

Calibration Successful

 will display for a few seconds to indicate a successful 

calibration and then the instrument will return to the run screen. 

8. 

If the calibration is unsuccessful, an error message will display on the screen.  
Press the 

Cal

 key to exit the error message and return to the run screen.