Pro
2030
Quick Start Guide
Item #605057
This Quick Start Guide is meant to serve as a quick reference in calibrating and
operating the Pro2030. It is not intended to replace the information found in the Users
Manual.
Setting Dissolved Oxygen Sensor & Membrane Type
The instrument’s dissolved oxygen (DO) sensor type must be configured for the sensor
installed. Failure to do this may result in damage not covered under warranty. The
correct DO membrane type must be configure for the membrane in use or the DO
readings will be erroneous.
The instrument will step you through an initial configuration when powered on for the
first time. This allows you to set the language, sensor and membrane options. Use the
up or down arrow keys to highlight the appropriate language, sensor, and membrane,
then press enter to confirm. Check the label on the box of the sensor and membrane
to determine which option you have. To change the sensor or membrane type after the
initial configuration, press the
Menu
key.
Installing the DO Sensor
Prior to use, the dissolved oxygen (DO) sensor must be installed on the Pro2030 cable.
Failure to install the DO sensor make cause permanent damage to the cable not
covered under warranty.
1.
Remove the red plastic plug from the cable’s port by pulling it straight out of the
port. This can be discarded.
2.
Remove the red plastic plug from the DO sensor’s connector by pulling it
straight off the sensor. This can be discarded.
3.
Ensure both the sensor connector and cable connector are clean and dry.
4.
Grasp the sensor with one hand and the cable bulkhead in the other.
5.
Push the sensor into the connector on the cable until it is firmly seated and only
1 o-ring is visible. Failure to properly seat the sensor may result in damage.
6.
Twist the sensor clockwise to engage the threads and finger tighten.
Do NOT
use a tool
. This connection is water-tight.