4. 

Press 

Enter

 to complete the calibration. Or, press 

Cal

 to cancel the calibration 

and return to the Run screen.    

5. 

Calibration Successful

’ will display for a few seconds to indicate a successful 

calibration and then the instrument will return to the Run screen.

6. 

If the calibration is unsuccessful, an error message will display on the screen. 
Press the 

Cal

 key to exit the calibration error message and return to the Run 

screen.  See the Troubleshooting guide for possible solutions.

Taking Measurements

1. 

Before taking measurements, be sure the instrument has been calibrated to 
ensure the most accurate readings.  

2. 

Turn the instrument on and wait 5-15 minutes if using a polarographic sensor.

3. 

Install the sensor guard and then insert the probe into the sample.

4. 

Ensure the conductivity sensor is completely submerged in the sample. The two 
holes near the cable should be covered by the sample for accurate conductivity 
readings.  

5. 

Move the probe in the sample at a rate of at least 6 inches (16 cm) per second if 
using the yellow membrane and 3 inches (7.62 cm) per second if using the blue 
membrane.

6. 

While continuing to provide sample movement, wait for the readings to 
stabilize. 

7. 

Highlight 

Save

 and press 

Enter

 to store the reading. The instrument will confirm 

that the reading was successfully saved.