Visit
YSI.com/Pro2030
to find the
User Manual, specs, and accessories.
Item# 605057 Dwg# 605057
November 2019 Rev C
1725 Brannum Lane
Yellow Springs, OH 45387
(800) 765-4974, (937) 767-7241
info@ysi.com
4.
Press
Enter
to complete the calibration. Or, press
Cal
to cancel the calibration
and return to the Run screen.
5.
‘
Calibration Successful
’ will display for a few seconds to indicate a successful
calibration and then the instrument will return to the Run screen.
6.
If the calibration is unsuccessful, an error message will display on the screen.
Press the
Cal
key to exit the calibration error message and return to the Run
screen. See the Troubleshooting guide for possible solutions.
Taking Measurements
1.
Before taking measurements, be sure the instrument has been calibrated to
ensure the most accurate readings.
2.
Turn the instrument on and wait 5-15 minutes if using a polarographic sensor.
3.
Install the sensor guard and then insert the probe into the sample.
4.
Ensure the conductivity sensor is completely submerged in the sample. The two
holes near the cable should be covered by the sample for accurate conductivity
readings.
5.
Move the probe in the sample at a rate of at least 6 inches (16 cm) per second if
using the yellow membrane and 3 inches (7.62 cm) per second if using the blue
membrane.
6.
While continuing to provide sample movement, wait for the readings to
stabilize.
7.
Highlight
Save
and press
Enter
to store the reading. The instrument will confirm
that the reading was successfully saved.