Warranty and Reordering Information

We guarantee that the ProDSS battery pack will be free from defects in materials and workmanship at the date of 
purchase for up to one year. Replacement decisions are at the sole discretion of YSI. This does not cover cosmetic 
damage or damage due to accident, misuse, abuse, negligence, commercial use, or modification of or to any part of 
the battery. YSI does not inspect or repair batteries. 

Do not ship batteries to YSI unless instructed to do otherwise.

A replacement battery (item #626846) can be ordered by calling 

(937) 767-7241

 or our toll free number at 

800-765-

4974

 and speaking with Customer Service or by sending an email to 

orders@ysi.com

 

 

Battery Replacement

Note: New instruments come with a pre-installed battery. Perform the following procedure when replacing the battery.

 

1. Remove battery cover 

The battery cover is located on the back of the ProDSS. To remove the battery 
cover, unscrew (counter-clockwise) the four screws. 

 

Note: The battery cover screws are captive and cannot be removed. 

2. Remove old Li-Ion battery pack and gasket/cradle 

With two fingers, grasp the black connector that connects the battery to the 
instrument. Pull the connector straight up to disconnect. Once disconnected, 
remove the battery pack and the battery gasket/cradle.  
 
A new gasket/cradle is included with each battery pack and should replace 
the old gasket/cradle to prevent water leaking into the instrument casing. 

 

3. Insert new Li-Ion battery pack and gasket/cradle

Inspect the new battery pack and battery gasket/cradle for damage. Please 
contact YSI customer service if the new battery pack and/or replacement 
gasket is damaged. 

Carefully insert the new gasket and ensure it is seated properly. Align the 
battery pack connector wire terminals with the three instrument pins, then 
connect the battery pack to the instrument.   

4. Replace battery cover

Ensure the battery gasket/cradle is properly seated and carefully reinstall the 
battery cover. 

     CAUTION:

 

Do not over-tighten the cover screws. This may cause damage 

to the instrument and/or cover which may allow water to leak into the 

battery compartment. 

Charging the instrument after replacing the battery is not required. New 
batteries are shipped at less than 50% full capacity. To charge the instrument, 
use the AC power adapter, a computer USB port, or an external USB battery 
pack. Always charge the instrument in an open area away from flammable 
materials, liquids and surfaces. 

DO NOT

 charge or handle batteries that are 

hot to the touch.

 

item# 626979REF dwg# 626979 

July 2016 Rev B 

YSI.com

!