31
View Calibration Records
From the Files Menu, highlight
View Calibration Records
and press enter. Select the channel from which you’d like to view the 10
most recent calibration records. Options include: DO, Conductivity, ISE1, ISE2, and Barometer. Use the arrow keys to scroll through
the records.
Calibration record content varies based on the sensor type.
Conductivity Sensor Calibration Record Content
• Date and Time
• Method (Spec Cond, Cond, Salinity)
• Cal Value (value of calibration solution)
• Sensor Value
• Temperature Reference (User selected in Sensor Setup menu)
• Temperature Compensation Coefficient %/°C (User selected in Sensor
Setup menu)
• TDS Constant (User selected in Sensor Setup menu)
• Temperature
• Cal Cell Constant
• Calibrate Status
DO Sensor Calibration Record Content
• Date and Time
• Method (%, mg/L)
• Cal Value
• Sensor Value (Sensor Current)
• Sensor Type (Polarographic/Galvanic)
• Membrane Type (Teflon Black, PE Yellow, PE Blue)
• Salinity Mode (user entered value if in Manual Salinity Mode)
• Temperature
• Barometer
• Calibrate Status
pH Sensor Calibration Record Content – Up to 6 Calibration Points
• Date and Time
• Buffer Value
• Sensor Value (mV)
• Temperature
• Slope (mV/pH)
• Slope (% of ideal)
• Calibrate Status
Figure 50
View Calibration
Records
03/03/20 10:34:42 AM
View Cal. Records
100%
Conductivity [10 of 10]
Date: 03/03/20
Time: 07:18:35AM]
Method: Salinity
Cal Value:
45.38 SAL ppt
Sensor V.:
45.38 SAL ppt
Temp. Ref.: 25.0 °C
Temp. Coeff.: 0.0191 %/°C
TDS Constant: 0.65
Show Data