background image

7

Ⅳ . SETTING OF PRESSURE

1. Switch contact confi guration

・  Composition  of  contact  of  this  instrument  is  as  shown  on 

the right.

・  If  the  instrument  is  not  applied  by  differential  pressure  we 

call such “

normal condition

” .

・  At  normal  condition,  the  contact  between  COM.  and  N.C. 

is closed, and between COM. and N.O. is open.

・  With  increase  of  differential  pressure  and  reaching  the  actuation  pressure,  the  contact  will  be  changed 

over ; between COM. and  N.C. is opened and between COM. and N.O. is closed.

2. Setting of upper limit/ lower limit

・  When  differential  pressure  applied  to  the  instrument  is 

increased  starting  from  zero  and  changes  over  the  electric 
contact from N.C. (normal close) to N.O. (normal open), the 
pressure at this moment is called “

actuation pressure

” .

・  When this differential pressure decreases from the pressure 

higher  than  this  actuation  pressure  activating  electrical 
contact and returning the contact from N.O. (normal open) 
to N.C. (normal close), the pressure at this moment is called 

reset pressure

” .

・  There is certain difference between actuation pressure and reset pressure and this is called “

dead band

” .

・  There are two types of the instrument. One is called “

upper limit setting type

” type (H) whose scale of 

setting knob is adjusted on actuation pressure,  the other is called “

lower limit setting type

” type (L) 

which is adjusted on reset pressure.

・  Be  careful  about  the  difference  between  the  upper  and  lower  limit  setting  switched  and  also  understand 

the relation between the pressure increase/ decrease and the contact composition before designing the 
operational sequence.

Depending  on  the  purpose  of  use,  choose  the  instrument 
either 

“upper limit setting type”

or

 “lower limit setting type”

Difference of contact movement between
“upper  limit  setting  type”  and  “lower  limit 
setting type”

In case of 1-6 kPa range with set value 2 kPa. 
(dead band is 0.6 kPa)

Adjusting of scale

Name plate

Upper limit 

setting type

Adjusted in 

activating pressure

In case of “upper limit setting type”

When the differential pressure increases, the 
circuit  is  closed  between  N.O.  and  COM.  at 
2 kPa.
Then  as  the  differential  pressure  decreases, 
the  circuit  is  opened  between  N.O.  and 
COM. at 1.4 kPa.

Lower limit 

setting type

Adjusted in 

reset pressure

In case of “lower limit setting type”

When the differential pressure increases, the 
circuit  is  closed  between  N.O.  and  COM.  at 
2.6 kPa.
Then  as  the  differential  pressure  decreases, 
the  circuit  is  opened  between  N.O.  and 
COM. at 2 kPa.

The scale of the setting knob is adjusted for both upper and lower limit setting type at our factory before 
shipping. It is not allowed for customers to convert from the upper limit setting to the lower limit setting or 
vice versa.

ᵢᶇᶄᶄᶃᶐᶃᶌᶒᶇᵿᶊᴾᶎᶐᶃᶑᶑᶓᶐᶃ

ᵟᶁᶒᶓᵿᶒᶇᶍᶌᴾᶎᶐᶃᶑᶑᶓᶐᶃ

ᵰᶃᶑᶃᶒᴾᶎᶐᶃᶑᶑᶓᶐᶃ

ᵱᶕᶇᶒᶁᶆᶇᶌᶅᴾᶂᶇᶄᶄᶃᶐᶃᶌᶁᶃ
ᵆᶂᶃᵿᶂᴾᶀᵿᶌᶂᵇ

ᵲᶇᶋᶃ

㻱㼤㼜㼘㼍㼚㼍㼠㼕㼛㼚㻌㼐㼕㼍㼓㼞㼍㼙㻌㼛㼒

㼟㼣㼕㼠㼏㼔㼕㼚㼓㻌㼐㼕㼒㼒㼑㼞㼑㼚㼏㼑㻌㼜㼞㼑㼟㼟㼡㼞㼑

ᾝᵌᾒᵌ

ᾒᾞᾜᵌ

ᾝᵌᾞᵌ

ᵧᶌᶁᶐᶃᵿᶑᶃᴾᶍᶄ
ᶂᶇᶄᶄᶃᶐᶃᶌᶒᶇᵿᶊ
ᶎᶐᶃᶑᶑᶓᶐᶃ

Summary of Contents for MS65

Page 1: ... U C T I O N M A N U A L 653 0031 1 2 3 Nishi shiriike cho Nagata ku Kobe Japan TEL 078 631 6000 FAX 078 631 6020 READ THE INSTRUCTION MANUAL MANOSTAR SWITCH MS65 MS65F WITH AIR FILTER TR MS65 E09 RoHS Compliant ...

Page 2: ...Caution of piping 6 7 Installation position 6 Ⅳ SETTING OF PRESSURE 1 Switch contact configuration 7 2 Setting of upper limit lower limit 7 3 Reset time 8 Ⅴ ABOUT BUILT IN MICRO SWITCH 1 Material of contact 8 2 Contact material and applicable mecanism for low voltage and low current circuit 8 3 Protection of contact of switch 9 4 About measured gas and contact failure 10 Ⅵ ACCESSORIES 1 Accessory f...

Page 3: ... instrument The instrument is measurable for air and non corrosive gas only The machine is exclusive use of dry air 90 RH or less Using measuring the water or oil it may be damaged and causes the accident Avoid using where the instrument is subject to many vibration and impact Using the instrument where intensified vibration and impact may be damaged the instrument It is expected that gas leaks of ...

Page 4: ... Pa 1 2E 0 2 1 2 0 030 kPa 0 120 kPa 3E 0 5 3 0 080 kPa 0 400 kPa 6E 1 6 0 15 kPa 0 60 kPa Silicone rubber 12E 2 12 0 30 kPa 1 20 kPa 30E 5 30 0 75 kPa 4 00 kPa Out of standard 㹋㹑㸴㸳 㹄 㹆 㹔㹊 㸱㸮㸮 㹂 㸫㹉 㸫㹉 㹂 䠬䠽 㹃 䡇䠬䠽 㹆 㹊 㹔 㹋 㹋㹊 㹔㹊 㻲㼛㼞㻌㼓㼑㼚㼑㼞㼍㼘㻌㼑㼘㼑㼏㼠㼞㼕㼏㼍㼘㻌㼘㼛㼍㼐 㻲㼛㼞㻌㼘㼛㼣㻌㼑㼘㼑㼏㼠㼞㼕㼏㼍㼘㻌㼘㼛㼍㼐 㼂㼀㻌㼟㼠㼞㼍㼕㼓㼔㼠㻌㼠㼥㼜㼑㻌 㼂䠨㻌㼞㼛㼠㼍㼞㼥㻌㼘㼛㼛㼟㼑㻌㼠㼥㼜㼑 㻹㼀㻌㼠㼟㼠㼞㼍㼕㼓㼔㼠㻌㼠㼥㼜㼑 㻹䠨㻌㼞㼛㼠㼍㼞㼥㻌㼘㼛㼛㼟㼑㻌㼠㼥㼜㼑 㻲㼛㼞㻌㼡㼜㼜㼑㼞㻌㼘㼕㼙㼕㼠㻌㼟㼑㼠㼠㼕㼚㼓 㻲㼛㼞㻌㼘㼛㼣㼑㼞㻌㼘㼕㼙㼕㼠㻌㼟㼑㼠㼠...

Page 5: ...de 10 10 to 50 Hz Acceleration 39 m s2 each two hours on triaxial directions 100 m s2 each six times on triaxial direction Between each terminal and case 20 MΩ or more 500 V DC megger Between each terminal and case 1500 V AC 50 60 Hz for one minute Phenol resin and aluminum die cast 1 Vinyl plastic or rubber tube I D 6 Piping applicable model attached with connector for vinyl or plastic tube 2 Met...

Page 6: ...ectrical load type The general electrical load type is decreased contact reliability refer to p 8 The withstanding pressure of the pressure receiving element means The max pressure under which the diaphragm withstands distortion or breakage withstand single pressure The single pressure acting on either H or L side The withstanding pressure of the instrument body means The max pressure under which ...

Page 7: ...de piping connector blue or L The high pressure port red or H should be opened to atmosphere but do not remove the piping connector c Measurement of differential pressure Connect the tube from the high pressure piping connector to the high pressure port red or H and from the low pressure piping connector to the low pressure port blue or L 4 Connectors and piping material a Connectors for vinyl or ...

Page 8: ...nnect it with the pressure receiving instrument through a metal tube which is long enough to cool down the high temperature gas d Errors caused by long distance piping The speed of response is delayed when the product is used for remote monitoring In such application the I D of the connection tube should be as large as possible The time constant is almost inversely proportional to the inner cross ...

Page 9: ...e Be careful about the difference between the upper and lower limit setting switched and also understand the relation between the pressure increase decrease and the contact composition before designing the operational sequence Depending on the purpose of use choose the instrument either upper limit setting type or lower limit setting type Difference of contact movement between upper limit setting ...

Page 10: ... small rivet contact point 2 Contact material and applicable mecanism for low voltage and low current circuit The schematic diagram in the right explains the proper range of load voltage and current for which silver or gold alloy contact is generally used when the contact is used for opening or closing the low voltage and low current circuit with a micro switch When referring to the diagram keep t...

Page 11: ...iode and avoid to apply high voltage to inbetween the contacts This method makes the return time of inductive load slower Choose the diode rated current is more than load current and counter withstand voltage more than 10 times of power supply voltage Diode Zener diode method Diode method proves to be effective when the return time takes too long for inductive load such as relay When you choose ze...

Page 12: ...ic acid and it may also cause electrical contact failure or malfunction of switches b Silicone gas If the gas to be measured contains silicon produced from silicon contained in oil grease filling agent and so on the gas produced by arc generates by opening closing the contact accumulates on the contact surface and causes contact failure Take corrective measures such as the removal of the gas sourc...

Page 13: ...again to fix Replacement of air filter In case of using model with air filter MS65F depending on amount of work conduct disassembly and clean out once in 2 years periodically and replace filters and packings with new ones Filters are constructed as shown in Fig 1 How to change 1 Loosen and take out four assembly screws a from the instrument body and disassemble filter unit 2 Remove the used filters and ...

Page 14: ...G VT6 S Product code FG MT6 S T joint For vinyl or plastic tube Product code TVT6 4 6 For branching different I D 6 and 4 Product code TVT6 6 6 For branching the same I D 6 and 6 Product code Applicable range max value For vinyl or plastic tube For metal tube Pa kPa RS VT6 02 RS MT6 02 200 Pa or less RS VT6 03 RS MT6 03 300 500 1000 Pa RS VT6 04 RS MT6 04 2 3 5 10 kPa RS VT6 06 RS MT6 06 20 kPa or...

Page 15: ...talog specifications or manual Handling or usage of the product other than as described in our product catalog specifications or manual 2 Breakdown due to a reason other than a fault within our product 3 Any product that has been modified or repaired by a party other than us 4 Any breakdown due to a reason that was not foreseeable based on scientific and technical standards applied at the time of ship...

Reviews: