4. Operating Procedures
Useful functions (Output terminal)
Before operating the unit
Be sure to follow instructions in this manual for operating the product. Operations other
than those specified in this manual may cause a trouble. Also take care that the warranty
may be void if any operation other than those specified in this manual is performed.
CAUTION
1. Be sure to turn the breaker OFF before making any connections.
2. Be sure to use the unit at or below the rated capacity when you use the alarm output
and the time-up output.
3. Be sure to connect a recorder having an input impedance of 600
Ω
or less to the
temperature output terminal.
4. Make any connection secure with the screws attached to the terminal block.
Connecting procedures
Securely connect to the terminal to be used.
The alarm output and the time-
up output are “ON” (relay contact closed) at the time of
outputting.
When temperature output is used, use a shielded cable to prevent noises.
Connection terminal
※
You cannot install the time-up output terminal and the alarm output terminal at the same time. Note
that the standard external alarm output terminal will be eliminated when you install the time-up
output terminal.
When you install all output terminals, two layouts 1 and 2 above are possible.
Layout 1
Layout 2