21
4. Operating Procedures
Operation steps
(
Quick Auto Stop operation
)
How to start quick
auto stop operation
This operation mode is used to when you
“want to stop Fixed Temp
operation after several hours automatically in the middle of it
”. Quick
Auto Stop operation is a function to allow auto stop timer setting
during operation.
MEASURED TEMP.
℃
HEATER
ENTER
SET
TEMP.
OVER TEMP .
RUN
STOP
SUB
MENU
TIMER
FIXED
TEMP.
PROGRAM
PROTECTOR
ALARM
AUTO STOP
AUTO START
FIXED TEMP.
PROGRAM
1.Set time period before stop during Fixed Temp operation
Make sure that the Fixed Temp operation lamp is on indicating the
unit is in the fixed temp operation mode.
Press
TIMER
key.
The characters tim
that means a timer is displayed on the
measured temperature indicator and the current set period of time
flashes on the set temperature indicator.
Use
▽
and
△
keys to set a period you want.
About the timer function
Maximum settable time for the timer is 999 hours 50 minutes.
Time can be set in the unit of minutes up to 99 hours 59 minutes.
Time period 100 hours and longer shall be set in the unit of 10
minutes.
You can continuously change set time and change to the setting you
want quickly by keeping
▽
and
△
keys pressed. Press
▽
and
△
keys once at a time to fine adjust a time.
MEASURED TEMP.
℃
HEATER
ENTER
SET
TEMP.
OVER TEMP .
RUN
STOP
SUB
MENU
TIMER
FIXED
TEMP.
PROGRAM
PROTECTOR
ALARM
AUTO STOP
AUTO START
FIXED TEMP.
PROGRAM
2. Start timer operation
When you have set a time you want, press
RUN/STOP
key for
about one second. Timer operation will start when the Fixed Temp
operation lamp and the Auto Stop lamp are lit. Timer starts when
RUN/STOP
key is pressed.
3. Stopping and ending Timer operation
Operation will stop automatically when the set time comes.
The buzzer sounds for about five seconds to tell you that operation
has stopped. At this time, the set temperature indicator display the
character End
that means completion of operation with the
Fixed Temp operation and the Auto Stop lamp are lit.
Press
RUN/STOP
key for about one second to end the Timer
operation mode. The screen will switch to the initial setting screen.
When you want to
correct
a
set
temperature or a set
time or check settings
When you want to change a set temperature during operation, keep
FIXED TEMP.
key pressed to enter the setting mode and change the
temperature. After changing the temperature, press
ENTER
key to
complete. When want to change a set temperature during operation,
keep
TIMER
key to enter the setting mode and change the time. At
this time, you need to set a time calculated by adding time period to
add to the time already passed. After changing the setting, press
RUN/STOP
key to complete.
Press
▽
key to display the set temperature, the operation mode, and
the remaining time on the set temperature indicator.
Summary of Contents for DKG610
Page 2: ......