33
In the Event of Failure
Trouble Shooting
Problem
Possible Cause
Solution
The device does not start
when turning on the power
switch.
Bad condition of earth leakage breaker
Replace the part
Bad condition of power switch
Problem in power source
Use an appropriate power supply.
Temperature does not rise.
Heater is disconnected
Replace the part
Thermal fuse is disconnected
Bad condition of thermo regulator
Takes too much time to
raise temperature.
Bad condition of temperature sensor
Replace the part
Heater is disconnected
Bad condition of thermo regulator
Problem in power source
Use an appropriate power supply.
Heater is not turned off after
reaching to preset value.
Bad condition of SSR
Replace the part
Bad condition of thermo regulator
Thermal fuse blows out.
Bad condition of fan
Replace the part
Bad condition of thermo regulator
When power failure occurs
…
When power is supplied after a power failure, the device automatically starts operation again with
the same state as just before the power failure. It is danger that the device starts unattached
operation after a power failure.
We recommend for you to turn off the switch of device if a power failure occurs during operation.
In the case if the error other than listed above occurred, turn off the power switch and primary
power source immediately. Contact the shop of your purchase or nearest Yamato Scientific
Service Office.