4. How to Use 

Replacement of fluorescent light 

Replacement of fluorescent light

Standard

Desk-top

Low-ceiling types

 

 

Remove  the  decorative  panel  (upper).  It  can  be  removed  by  removing  the  screws 

attaching the panel, and by lifting the panel upward. 

 

Remove the screws and remove the lighting cover. There is a fluorescent light in it. 

 

Replace the fluorescent light with new one. 

 

Return to the original condition in the reversed order of removal as shown above. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Fig. 4-4 

Decorative panel 

  (upper) 

Fluorescent  temp 

Lighting  cover   

Screws 

Summary of Contents for LDS

Page 1: ...d To use this product correctly please read thoroughly this Instruction Manual and Warranty Card before using After reading please save them in a suitable place for future use Warning Please read care...

Page 2: ...t of fluorescent light 27 5 Handling Precautions 29 Precautions for operation 29 Working surface excluding the desk top type 29 Exhaust blower 30 Ambient environment 30 Maintenance 30 Others 30 6 How...

Page 3: ...and Health Law 39 10 Specifications 40 11 Wiring Diagram 42 12 List of Hazardous Materials 43 13 List of Spare Parts 44 Mechanical part 44 Electric part 44 14 Standard Installation Manual 45 15 Memor...

Page 4: ...sical damage electric shock bone fracture poisoning and others which have any aftereffect or require medical treatment by going to hospital for a long time or admitting to hospital Note 2 Minor injury...

Page 5: ...neral caution Electric shock Risk of burn Do not heat with no water Water leak Use water only Toxic substance Prohibition General inhibition Do not use fire Do not disassemble Do not touch Mandatory G...

Page 6: ...pheric temperature in the installation site must be maintained within 5 35 C Do not use this product for experiments of perchloric acids If a perchloric acid is used the exterior and interior of this...

Page 7: ...effect may reduce and the discharged gas may be flowing in the reversed direction which will have an adverse effect on human bodies Do not perform any experiments when the exhaust blower stops If the...

Page 8: ...cover are made of glass If a fluorinated acid is used in the hood the glasses may be corroded by the acid and become foggy translucent to make it difficult to see the inside clearly Do not use fire ne...

Page 9: ...arge amount of gases which have been generated during experiments adhere to the interior and working surface as fouling Clean with water periodically Working surface If you want to use a heat source s...

Page 10: ...rectly comes into the opening of fume hood or comes into contact with the front face and then goes down into the opening Near air inlet for air conditioning ventilation on the ceiling or on the wall T...

Page 11: ...type code of this product stands for LDS 150 S A 2 Damage In case you find any damages to the product when you received it please contact the transporter for confirmation as soon as possible Shape Non...

Page 12: ...ork shop if no appropriate person is available Operation without grounding may cause electric shock accidents Do not connect the grounding wire to gas plumbing grounding wire of telephone or lighting...

Page 13: ...witch and during operation which may cause fire or explosion Refer to P 43 Hazardous material list for flammable and explosive gases 6 Do not modify the product 7 Install on a level surface Do not mod...

Page 14: ...not use the power cord with it tied in a bundle Bundled power cord may be heated which may cause fire Do not modify or forcedly bend twist pull the power cord Otherwise fire or electric shock may be...

Page 15: ...on Next remove two blocks which are attached to the lower part of both side plates The blocks are secured with four M8 bolts from the bottom Then lift the hood and place on the frame Caution The hood...

Page 16: ...hrough hole in the right The other type are unnecessary Bring the power supply cord for power outlet and lighting and the cord for SW circuit board into the frame through holes of left side plate and...

Page 17: ...ace it on the protecting blanket and pull the blanket together with the hood to the installation position When installation is over pull out the protecting blanket Caution Five workers must take part...

Page 18: ...t board go through the hole on the right side plate into the frame for connection The inverter type ballast and SW circuit board are connected through connectors 2 5 Connect the primary power wiring t...

Page 19: ...ing piping is a flexible tube with it end threaded nominal diameter 1 2 B male or female Connect this end to a primary piping such as a riser from the floor Ask an authorized local company for connect...

Page 20: ...10 3 5 0 120 0 2 0 0 15 0 F LOW M ON IT OR LDS180 Fume Hood 1889 SINCE 6 3 1 550 6 1 5 6 0 68 5 3 2 4 8 10 12 11 1 9 15 13 16 18 7 21 24 22 23 19 30 17 20 6 27 5 28 29 26 25 14 3 Part Names and Functi...

Page 21: ...S 180SA Air curtain type 1200 150 550 750 320 1020 865 43 132 850 FLOW MONITOR L D S 1 8 0 F u m eH o o d 1889 SINCE 1800 2600 1750 210 615 60 885 3 1 4 2 10 11 1 15 13 16 18 5 6 17 7 21 24 22 23 26 2...

Page 22: ...Fume Hood 1889 SINCE 2 10 0 7 50 12 5 0 1 5 0 1 0 0 1 13 6 1 5 60 50 0 86 5 753 210 35 0 1 2 00 2 00 1 50 LDS180 Fume Hood 1889 SINC E 6 3 F LO W MO NI TO R 15 5 0 6 85 3 Part Names and Functions LDS...

Page 23: ...tain the windowpane in position in case of broken wire Use in accordance with the opening extent of windowpane 11 Exhaust blower activation monitor The movement of feathers visualizes the condition of...

Page 24: ...26 Drain hose Drain hose from discharge air duct 27 Upper section Hood and associated equipment 28 Lower section Provides connections to utilities and stowing space for the special purpose cart 29 Suc...

Page 25: ...erating It flashes when the delay timer is operating 3 Exhaust blower stop key OFF Used to stop the exhaust blower Generally the ventilator stops with a delay set by the incorporated timer for dischar...

Page 26: ...blower The supply blower stop switch on the side of distribution board enables stop of the supply blower only This can be used for loading unloading of the powder sample The windowpane can be opened c...

Page 27: ...jects such as pieces of paper or vinyl from being sucked If an object is adhered to the net stop the test temporarily and remove the object Then resume the test If an object is adhered to the net the...

Page 28: ...e on Prevention of Hazards Due to Specified Chemical Substances the 16 division measurement is mandatory On the other hand for applications other than above the wind speed at the center of each segmen...

Page 29: ...into five equal portions and measure the air flow in a center of each of these portions with a wind speed indicator Fig 4 3 Fig 4 3 From the average of air flow values obtained from five points calcu...

Page 30: ...removed by removing the screws attaching the panel and by lifting the panel upward Remove the screws and remove the lighting cover There is a fluorescent light in it Replace the fluorescent light wit...

Page 31: ...ling The fluorescent tube is inside the cover Change the tube with new one Reverse the removal procedure after change 4 6 Warning Wear protection glasses rubber gloves protection mask etc when perform...

Page 32: ...in the hood must be limited to the extent as shown in the table below Type Heat quantity used kcal hr Equivalent Gas burner Electric heat Front wind speed at 0 5 m s LDS 120 1300 1 1 5 kW LDS 150 2000...

Page 33: ...ire is 3 years but it may be shortened depending on the used condition Interior and working surface of fume hood must be periodically cleaned with water If the exhaust blower is corroded or whitened r...

Page 34: ...es and handle attachments on front window sash are made of ABS plastic When cleaning them use a mild detergent not organic solvents Avoid using organic solvents because they may cause damage to handle...

Page 35: ...of drain tube Periodically check the drain tube for drop out and blockage Check of spot sink Periodically check the spot sink for blockage and the bell trap for disconnection If the bell trap is disc...

Page 36: ...Major component Material Major component of body Exterior Steel plate epoxy polyester plastic paint stainless steel SUS304 Interior General purpose type Non asbestos special board fire proof melamine...

Page 37: ...ut not operating Defective motor Replace or repair the motor Open phase operation of motor Check that the power supply connection is single phase Check there is a discontinuity in the wiring Large noi...

Page 38: ...duced air flow caused by increased pressure loss due to built up of particles in the duct Clean the inside of duct to remove particles Outside air enters the duct because of broken fan duct or loosene...

Page 39: ...f purchase In accordance with the term of guarantee booklet the product in question shall be repaired or replaced at no charge Contact your dealer or Yamato Scientific s office if you want to repair y...

Page 40: ...formation consult your dealer or our office Summary of maintenance check items and intervals Category Maintenance check item Description Interval Performance Discharge function Front wind speed measur...

Page 41: ...packing 19 mm 1 yr 3 yr Water tap 1 yr 3 yr Insulator board 1 yr 3 yr Exhaust blower activation monitor 1 yr 3 yr Door drop prevention stopper 1 yr 3 yr Note 1 Each maintenance interval does not mean...

Page 42: ...ks by using the table below as a reference We also provide periodic self check services charged Consult your dealer or one of our company s offices if interested No Location Check Item 1 Hood 1 Condit...

Page 43: ...l plate SPCC chemical proof powder coating with adjuster Interior Non asbestos special board fire resistant melamine resin decorative Baffle plate Multi slit type Windowpane Temper glass 6 mm Vertical...

Page 44: ...h duct PVC with ID 400 mm OD 100 mm Connected with a half flange Applicable air supply duct PVC with ID 211 mm OD 216 mm External dimensions W D H mm 1200 750 1020 2600 1500 750 1020 2600 1800 750 102...

Page 45: ...Part name MCB1 Circuit breaker LED1 Exhaust blower lamp MCB2 Circuit breaker LED2 Fluorescent light MC1 Magnetic contactor CN Power outlet MC2 Magnetic contactor FL Fluorescent light THR1 Thermal rela...

Page 46: ...rchlorate sodium perchlorate ammonium perchlorate and other perchlorates 3 Potassium peroxide sodium peroxide barium peroxide and other inorganic peroxides 4 Potassium nitrate sodium nitrate ammonium...

Page 47: ...tific Refer to 3 Part Names and Functions P 17 for numbers Electric part Symbol Part name Specification Manufacturer MCB1 Breaker 200V 15A BBW315 Panasonic MCB2 Breaker 100V 20A BS2022 Panasonic CN Po...

Page 48: ...able items when installed to plugs and breakers 1 Before Use Always connect 11 Specifications Specification power supply P 9 P 4 0 2 Start operation Start the operation 4 How to Use P 23 3 Air quantit...

Page 49: ...periodic maintenance is mandatory to keep your fume hood in a good condition It is often the case that associate work vendors will become unknown in the process of time To facilitate maintenance activ...

Page 50: ...caused Notice The content of this manual may be changed without prior notice We will replace the manual if it has a manufacturing defect There are minor differences in illustrations of products and me...

Reviews: