8
Operation Method
Preparation for Operation
1. Setting of shaking method
The unit is set to the rotary shaking method at the time of shipping. Perform the following procedures in
order to set it to the reciprocating shaking method.
①
Remove the shaking table when it is on the unit. Remove the two shaking table fixing screws by turning
them counterclockwise, then hold the handle of shaking table and lift it to remove it.
②
Change the attaching positions of the four flat countersunk head screws which are fixing the shaking
table fixing table in referring to the drawing shown below. Move the flat countersunk head screws from
the positions A to the positions B by using a Philips screwdriver. At this time, screw the flat countersunk
head screws which are put into the holes B into the reciprocating motion axis of the main unit surely.
Fig.1 Setting of Reciprocating Shaking Method
2. Setting of shaking table
The shaking table is not the standard accessory. Prepare it separately.
At the time of setting of the shaking table, hold the handle of the shaking table, then set it between the
front pins and the pusher pins (2 pieces) on the rear side of the shaking table fixing table by pushing the
rear side pins. Finally, set the shaking table fixing screws (2 pieces) to the hole positions, then tighten
them by your fingers.
Fig.2 Setting of Shaking Table
A: Screw hole for the rotary method setting.
B: Screw hole for the reciprocating method setting
Rotary motion plate
(For rotary method)
Reciprocating motion axis
(For rotary method)
Flat countersunk
Head screw(four pieces)
Shaking table fixing table
Shaking table
(non-standard accessory)
Fixing screws
Pusher pins
Shaking table fixing table
Setting pins
Summary of Contents for MK201D