background image

 

 

Operation Method 

Preparation for Operation 

 

1.  Setting of shaking method 

The unit is set to the rotary shaking method at the time of shipping. Perform the following procedures in 
order to set it to the reciprocating shaking method.   

  Remove the shaking table when it is on the unit. Remove the two shaking table fixing screws by turning 

them counterclockwise, then hold the handle of shaking table and lift it to remove it. 

  Change the attaching positions of the four flat countersunk head screws which are fixing the shaking 

table fixing table in referring to the drawing shown below. Move the flat countersunk head screws from 
the positions A to the positions B by using a Philips screwdriver. At this time, screw the flat countersunk 
head screws which are put into the holes B into the reciprocating motion axis of the main unit surely.   

 

 

 

Fig.1 Setting of Reciprocating Shaking Method 

 

2. Setting of shaking table 

The shaking table is not the standard accessory. Prepare it separately. 
At the time of setting of the shaking table, hold the handle of the shaking table, then set it between the 
front pins and the pusher pins (2 pieces) on the rear side of the shaking table fixing table by pushing the 
rear side pins. Finally, set the shaking table fixing screws (2 pieces) to the hole positions, then tighten 
them by your fingers. 

 
 

 

 

Fig.2 Setting of Shaking Table

 

 

 

A: Screw hole for the rotary method setting. 
B: Screw hole for the reciprocating method setting 

Rotary motion plate 

(For rotary method) 

Reciprocating motion axis 

(For rotary method) 

Flat countersunk 
Head screw(four pieces) 

Shaking table fixing table 

Shaking table 
(non-standard accessory) 

 

Fixing screws 

Pusher pins 

Shaking table fixing table 

Setting pins 

Summary of Contents for MK201D

Page 1: ...rchasing Shaking Device MK201D of Yamato Scientific Co Ltd To use this unit properly read this Instruction Manual thoroughly before using this unit Keep this instruction manual around this unit for referring at anytime WARNING Carefully read and thoroughly understand the important warning items described in this manual before using this unit ...

Page 2: ......

Page 3: ...ion 6 Component Names Functions 7 Operation Method 8 Preparation for Operation 8 Operation 9 Timer Operation method 9 Timer operation method 10 Handling Precautions 11 Maintenance Method 13 Daily Inspection and Maintenance 13 When not using this unit for long term When disposing 13 In the Event of Failure 14 Trouble Shooting 14 In such cases 14 After Service and Warranty 15 Specification 16 Wiring...

Page 4: ...f a problem that may cause a serious accident or ev en f atality CAUTION If the caution is ignored there is the danger of a problem that may cause injury damage to property or the unit itself Meaning of Symbols This sy mbol indicates items that urge the warning including the caution A detailed warning message is shown adjacent to the sy mbol This sy mbol indicates items that are strictly prohibite...

Page 5: ...aution electrical shock Caution scald Caution no road heating Caution not to drench Caution water only Caution deadly poison Prohibit Prohibit generally Prohibit inflammable Prohibit to disassemble Prohibit to touch Compulsion Compulsion generally Compulsion connect to the grounding terminal Compulsion install on a flat surface Compulsion disconnect the power plug Compulsion periodical inspection ...

Page 6: ...shock may result Never perform repair work yourself since it is dangerous and not recommended Do not use the power cord if it is bundled or tangled Do not use the power cord if it is bundled or tangled If it is used in this manner it can overheat and fire may be caused Do not process bend wring or stretch the power cord forcibly Do not process bend wring or stretch the power cord forcibly Fire or ...

Page 7: ...rounding wire green from ground adapter to a ground terminal When no grounding terminal is found Grounding to Electrical Equipment Technical Standards Section 19 class D Grounding Resistance Max 100Ω is required in Japan Contact a local dealer electrician or Yamato Sales office for location specific electrical requirements 2 Choose a proper place for installation Do not install this unit in a plac...

Page 8: ...c may be generated when the power switch is turned ON or OFF and fire explosion may result 4 Do not modify Modification of this unit is strictly prohibited This could cause a failure 5 Installation on horizontal surface Set this unit to the flattest place Not setting this unit with its 4 legs covered with rubber contacted to the setting place surface evenly could cause the vibration or noise or ca...

Page 9: ...ation To prevent take measures that putting up a notice of operating making an enclosure etc 8 Handling of power code Do not entangle the power cord This will cause overheating and possibly a fire Do not bend or twist the power cord or apply excessive tension to it This may cause a fire and electrical shock Do not lay the power cord under a desk or chair and do not allow it to be pinched in order ...

Page 10: ...ER indicator illuminates green Speed knob Controls mixing speed Turn clock wise to increase speed Speed indicator The screen displays the shaking speed RPM Timer Select of timer operating mode Refer to page 9 Timer operation method Fuse The fuse cut off the power supply in case of overcurrent occurrence In that case because exchange is necessary I take off a cap and let a fuse of the inside and pl...

Page 11: ...ich are put into the holes B into the reciprocating motion axis of the main unit surely Fig 1 Setting of Reciprocating Shaking Method 2 Setting of shaking table The shaking table is not the standard accessory Prepare it separately At the time of setting of the shaking table hold the handle of the shaking table then set it between the front pins and the pusher pins 2 pieces on the rear side of the ...

Page 12: ...ime of setting of the shaking speed When the power switch is set to TIMER it start timer operation When the unit is operating the speed indicator show the concussion speed About the setting method of the timer refer to next page Timer operation method When the shaking is completed turn the Speed knob counterclockwise to MIN then turn OFF the Power switch Timer Operation Time unit and time range mu...

Page 13: ...1 2 3 12 30 is selected with the time range selector at the lower left corner of the front panel and the selected time range appears in the window at the lower right part within the plastic frame of the time setting knob 3 Set Time Set time with time setting knob Timer set range will be 0 5 seconds to 100 hour 3 three Selectors The selectors can be turned clockwise and counterclockwise to select t...

Page 14: ...ropriately at the shipment from factory If the power voltage fluctuates widely or if the motor temperature rises abnormally the speed could be changed more or less If any abnormality including abnormal vibration and shaking main unit is detected reduce the speed by means of the speed adjusting knob If max Specified revolution is exceeded the abnormal vibration or noises could be caused and thus be...

Page 15: ...ing speed too quickly may send the agitator off center and into the sides of the container possibly causing damage or breakage During a thunder storm During a thunderstorm turn off the power key immediately then turn off the circuit breaker and the main power If this procedure is not followed fire or electrical shock may be caused Operation method Operate this unit with the method specified by thi...

Page 16: ...of reach of children Remove the driving parts Treat as large trash Environmental protection should be considered We request you to disassemble this unit as possible and recycle the reusable parts considering to the environmental protection The feature components of this unit and materials used are listed below Component Name Material Exterior Parts Frame Steel plate baked with melamine resin coati...

Page 17: ... a different site Speed is uncontrollable Speed controller is defective Motor is defective Replace Replace In such cases Symptom Check The unit will not operate even though power is turned on If the power cord is connected to an outlet securely If a power failure has occurred If the fuse has blown Shaking speed fluctuates during operation If the amount of specimen is too much If specimen is concen...

Page 18: ...ales office Check following items before contact Model Name of Product Production Number Purchase Date About Trouble in detail as possible Minimum Retention Period of Performance Parts for Repair The minimum retention period of performance parts for repair of this unit is 7 years after discontinuance of this unit The performance part for repair is the part that is required to maintain this unit Se...

Page 19: ...Shaking times 20 200rpm Continuous adjustment Shaking indicator Digital display Timer 0 5sec to 100hour A unit and scale is switched Power supply 50 60Hz AC115V 2A External dimensions W442 D415 H130mm Dimensions do not include protruding components Weight Approx 19kg Accessories Fuse 5A Shaking table Options Refer to a catalog of the YAMATO ...

Page 20: ...G 4IK25RGN A 4GN3RA AC200VAC 6μF TLNSRS AS01 NSRS AS01A NSRS RS01A NSRS FS01A C 87A NH04EH450EH 5A 6φ Name Power switch Timer Power Lamp Motor Gear Condenser Transformer Digital display control board Digital display board Interrupter board Motor control board Cable Fuse Symbol SW T PL M C Tr T 1 T 2 T 3 T 4 FUSE AC115V 50 60Hz ...

Page 21: ...mmonium perchlorate and other perchlorate Potassium peroxide Sodium peroxide Barium peroxide and other inorganic peroxide Potassium nitrate Sodium nitrate Ammonium nitrate and other nitrate Sodium chlorite and other chlorites Calcium hypochlorite and other hypochlorites INFLAMMABLE LIQUID Ethyl ether Gasoline Acetaldehyde Propylene chloride Carbon disulfide and other flammable substances having a ...

Page 22: ...r conduct what this document forbids Unexpected accidents or breakdown may result in Note The contents of this document may be changed in future without notice Any books with missing pages or disorderly binding may be replaced Instruction Manual for Shaking Device MK201D First Edition February 18 2016 Yamato Scientific America Inc 925 Walsh Avenue Santa Clara CA 95050 U S A http yamato usa com Tel...

Reviews: