YCO-1049-02-03
PPC-200W
Technical Manual
8
IV Parameter Lists
IV.B System Parameters
IV.C Factory Parameters
Key Std Function Description
Word Value
30 018 System ID 018 Do not change.
31 010 Gravity compensation 000 No compensation.
001~016 Compensation for a specifi ed area (Japan only.)
017~150 Acceleration due to gravity formula, (mm/s2 )-9700.
32 000 Linearity compensation 000 No compensation.
001~127 Positive compensation.
128~255 Negative compensation.
33 004 Steady-state sampling count 000~255
34 012 Steady-state count 000~255
35 008 Polarity steady-state count 000~255
36 020 Steady-state collapse count 000~255
37 004 Steady-state average count 000~255
38 010 Zero tracking timing 000 Disable zero tracking
001~255 Zero tracking at specifi ed counts
39 001 Neglectable change count 000~010 Number of counts
40 000 Simple test mode 000 User parameter valid
001 User parameter invalid
002 User parameter and simple test mode valid
Key Function Description
Word
51 Scale mode 000 Fixed single increment.
001 Multi-increments.
002 Do not use.
003 Do not use.
004 Change-over / gram mode - decimal pound mode.
005 Change-over / gram mode - lb/oz mode.
006 lb/oz mode.
007 Change-over / gram mode - continuous ounce mode.
52 Multi-increments, 000 Fixed single increment.
complex increment mode 001 Fixed accuracy, 3 increments.
002 Fixed accuracy, 2 increments.
003 Increment change at 50% full scale, 2 increments.
004 Increment change at 80% full scale, 2 increments.
005 Increment change at 64% full scale, 2 increments.
006 Increment change at 40% full scale, 2 increments.
53 Weighing capacity base value 000~099
54 Weighing capacity index value 001~004
55 Increment for lighter range 000 1
001 2
002 5
003 10
004 20
005 50
006 100
007 200