YCO-1049-02-03
PPC-200W
Technical Manual
IV Parameter Lists
IV.C Factory Parameters
10
Key Function Description
Word
71 Tare function 000 No tare function.
001 One-time tare.
002 Consecutive tare.
72 Zero reset under tare operation 000 Zero reset enabled.
001 Zero reset disabled.
73 Power off with ON key 000 ON key disabled for power off.
001 Power off after pressing ON key for two seconds.
74 Special function for stability 000 No function.
001 Enforcement.
002 EU mode.
75 Over - under feature 000 Disabled.
001 Enabled.
80 Internal resolution 000 Do not use.
001~255
81 1 timer (ms) 000 Do not use.
001~255
82 3 timer (ms) 000 Do not use.
001~255
91 Mechanical zero 1 000~255 Automatically set at span adjustment.
92 Mechanical zero 2 000~255 Automatically set at span adjustment.
93 Span coeffi cient 1 (small) 000~255 Automatically set at span adjustment.
94 Span coeffi cient 2 (small) 000~255 Automatically set at span adjustment.
95 Span coeffi cient 1 (large) 000~255 Automatically set at span adjustment.
96 Span Coeffi cient 2 (large) 000~255 Automatically set at span adjustment.
97 Regional number and gravity 000~155 Automatically set at span adjustment.
for span adjustment
99 Factory setting 000 Always displayed. * Do Not Change *