                Technical Manual

IV Parameter Lists

IV.C   Factory Parameters

10

    

Key                           Function                         Description

   

Word

      

71                         Tare function                       000  No tare function.

                                                                              001  One-time tare.
                                                                              002  Consecutive tare.
      72           Zero reset under tare operation        000  Zero reset enabled.
                                                                              001  Zero reset disabled.
      73                  Power off with ON key               000  ON key disabled for power off.
                                                                              001  Power off after pressing ON key for two seconds.
      74              Special function for stability           000  No function.
                                                                              001  Enforcement.
                                                                              002  EU mode.
      75                   Over - under feature                 000  Disabled.
                                                                              001  Enabled.
      80                     Internal resolution                   000  Do not use.
                                                                              001~255
      81                          1 timer (ms)                       000  Do not use.
                                                                              001~255
      82                          3 timer (ms)                       000  Do not use.
                                                                              001~255
      91                     Mechanical zero 1                  000~255   Automatically set at span adjustment.
      92                     Mechanical zero 2                  000~255   Automatically set at span adjustment.
      93               Span coeffi cient 1 (small)             000~255   Automatically set at span adjustment.
      94               Span coeffi cient 2 (small)             000~255   Automatically set at span adjustment.
      95                Span coeffi cient 1 (large)             000~255   Automatically set at span adjustment.
      96               Span Coeffi cient 2 (large)            000~255   Automatically set at span adjustment.
      97             Regional number and gravity          000~155   Automatically set at span adjustment.
                              for span adjustment                
      99                        Factory setting                     000  Always displayed.  * Do Not Change *

Summary of Contents for PPC-200W

Page 1: ...YCO 1049 02 03 YAMATO CORPORATION 1775 S Murray Blvd Colorado Springs CO 80916 USA Tel 719 591 1500 Fax 719 591 1045 YAMATO TECH CORPORATION 112 19425 Langley By Pass Surrey B C V3S 6K1 Canada Tel 604...

Page 2: ...ce the item to be weighed in the center of the platform Items placed on the edge of the platform may fall off and cause injury 5 When weighing a heavy large or unbalanced item make sure the item is st...

Page 3: ...only the AC adaptor supplied or an adaptor of the same output characteristics Capacity Increment Accuracy Maximum Tare 4 lb 2 kg 0 002 lb 0 001 kg 0 002 lb 0 001 kg 3 990 lb 1 995 kg 10 lb 4 kg 0 005...

Page 4: ...Turn the scale on and short the two TEST jumpers on the left of the PCB above the LCD 5 Once the scale is in Test Mode the display can be rotated through the following functions by pressing Internal C...

Page 5: ...d and it s standard value Example To select zero tracking time press eight times 4 User parameters are also set in the System Parameter Mode Press to cycle through system keywords 30 through 40 After...

Page 6: ...justment mode and the fraction i e 0 2 1 2 or 2 2 for three point of full scale in kilo grams that should be placed on the platform 3 If the scale is in two point span adjustment mode press to enter t...

Page 7: ...bration Mode II E Factory Parameter Mode 1 Enter Test Mode see section II A for details 2 Enter Calibration Mode see section II C for details 3 Press and to enter Factory Parameter Mode The steps are...

Page 8: ...rm Platform weight too high Platform weight too low Material between the platform and the scale body Weighing assembly in contact with wires or case Load cell connector loose or damaged Power board ma...

Page 9: ...erial interface 001 Speci ed commands acceptable 002 Sends zero detection 003 Sends motion detection 004 Sends continuously 04 000 Communication data 000 Net weight 1 batch 001 Net weight tare weight...

Page 10: ...9 001 Neglectable change count 000 010 Number of counts 40 000 Simple test mode 000 User parameter valid 001 User parameter invalid 002 User parameter and simple test mode valid Key Function Descripti...

Page 11: ...mode 000 0 0 00 001 0 0 0 002 0 0 003 004 Not used 63 Pound increment 000 008 64 Pound conversion coef cient 000 Conversion coef cient 1 00000 001 Conversion coef cient 0 705479 002 Conversion coef ci...

Page 12: ...01 Enabled 80 Internal resolution 000 Do not use 001 255 81 1 timer ms 000 Do not use 001 255 82 3 timer ms 000 Do not use 001 255 91 Mechanical zero 1 000 255 Automatically set at span adjustment 92...

Page 13: ...ent 003 002 003 003 65 Zero point range 019 019 019 019 66 Zero point range on plus side 012 012 012 012 67 Over scale 003 003 003 003 68 Display hold function 000 000 000 000 69 Hold release against...

Page 14: ...t 001 001 002 001 65 Zero point range 019 019 019 019 66 Zero point range on plus side 012 012 012 012 67 Over scale 003 003 003 003 68 Display hold function 000 000 000 000 69 Hold release against in...

Page 15: ...t 001 001 002 001 65 Zero point range 019 019 019 019 66 Zero point range on plus side 012 012 012 012 67 Over scale 003 003 003 003 68 Display hold function 000 000 000 000 69 Hold release against in...

Page 16: ...YCO 1049 02 03 PPC 200W Technical Manual V Wiring Diagram 14 Note The PPC 200W is not washdown with this option Note The PPC 200W is not washdown with this option...

Page 17: ...YCO 1049 02 03 PPC 200W Technical Manual VI Parts List 15...

Page 18: ...1 27 YAM 1249 200027 CPU Board Assembly 1 28 YAM 1249 200028 Platform Cover S S 1 29 YAM 1249 200029 Level Indicator 1 30 YAM 1249 200030 Sealing Screw 1 32 YAM 1249 200032 Upper Housing Assembly Inc...

Reviews: