YCO-1049-02-03

PPC-200W

                Technical Manual

I    Introduction

The PPC-200W is available in four capacities.  It displays units of either kilograms and one of the fol-

 low ing three: continuous pounds, continuous ounces, or pounds and ounces.  The specifi cations for 
each capacity are in the following table.

Additional specifi cations are provided in the operator’s manual.  It is important to note that various fac-

tory parameter settings depend upon the mode (continuous pound, continuous ounce, or pound and 
ounce) option selected when the scale was purchased.  See section IV.D for the standard set tings 
for each mode.

The PPC-200W is not washdown with the optional AC adaptor or the analog meter.
Remove the batteries when operating the scale with the AC adaptor.  Use only the AC adaptor supplied 

or an adaptor of the same output characteristics.

              Capacity                    Increment                            Accuracy                         Maximum Tare

            4 lb      2 kg          0.002 lb     0.001 kg            0.002 lb      0.001 kg            3.990 lb     1.995 kg
          10 lb      4 kg          0.005 lb     0.002 kg            0.005 lb      0.002 kg            9.975 lb     3.990 kg
          20 lb    10 kg            0.01 lb     0.005 kg              0.01 lb      0.005 kg            19.95 lb     9.975 kg
          40 lb    20 kg            0.02 lb       0.01 kg              0.02 lb        0.01 kg            39.90 lb     19.95 kg

1

II  Set-up Procedures

1.   Turn scale on.  Press 

  and 

  to enter User Pa ram e ter 

Mode.  The scale will indicate a keyword fol lowed by it’s value.

2.   To select the desired keyword, use the keys as follows:

           

  - increase keyword by one

           

  + 

  - decrease keyword by one

      Note:    Use the User Parameter List in Section VII to locate the 

function to be changed and it’s standard value.

     

      Example: To select lb/kg priority at power on press 

 

      nine times.

     

3.   To change a parameter value, use the keys as follows:

           

  - increase value by one

           

  + 

  - increase value by ten

           

  - decrease value by one

    II.A    User Mode

Keyword Parameter 

Value

Summary of Contents for PPC-200W

Page 1: ...YCO 1049 02 03 YAMATO CORPORATION 1775 S Murray Blvd Colorado Springs CO 80916 USA Tel 719 591 1500 Fax 719 591 1045 YAMATO TECH CORPORATION 112 19425 Langley By Pass Surrey B C V3S 6K1 Canada Tel 604...

Page 2: ...ce the item to be weighed in the center of the platform Items placed on the edge of the platform may fall off and cause injury 5 When weighing a heavy large or unbalanced item make sure the item is st...

Page 3: ...only the AC adaptor supplied or an adaptor of the same output characteristics Capacity Increment Accuracy Maximum Tare 4 lb 2 kg 0 002 lb 0 001 kg 0 002 lb 0 001 kg 3 990 lb 1 995 kg 10 lb 4 kg 0 005...

Page 4: ...Turn the scale on and short the two TEST jumpers on the left of the PCB above the LCD 5 Once the scale is in Test Mode the display can be rotated through the following functions by pressing Internal C...

Page 5: ...d and it s standard value Example To select zero tracking time press eight times 4 User parameters are also set in the System Parameter Mode Press to cycle through system keywords 30 through 40 After...

Page 6: ...justment mode and the fraction i e 0 2 1 2 or 2 2 for three point of full scale in kilo grams that should be placed on the platform 3 If the scale is in two point span adjustment mode press to enter t...

Page 7: ...bration Mode II E Factory Parameter Mode 1 Enter Test Mode see section II A for details 2 Enter Calibration Mode see section II C for details 3 Press and to enter Factory Parameter Mode The steps are...

Page 8: ...rm Platform weight too high Platform weight too low Material between the platform and the scale body Weighing assembly in contact with wires or case Load cell connector loose or damaged Power board ma...

Page 9: ...erial interface 001 Speci ed commands acceptable 002 Sends zero detection 003 Sends motion detection 004 Sends continuously 04 000 Communication data 000 Net weight 1 batch 001 Net weight tare weight...

Page 10: ...9 001 Neglectable change count 000 010 Number of counts 40 000 Simple test mode 000 User parameter valid 001 User parameter invalid 002 User parameter and simple test mode valid Key Function Descripti...

Page 11: ...mode 000 0 0 00 001 0 0 0 002 0 0 003 004 Not used 63 Pound increment 000 008 64 Pound conversion coef cient 000 Conversion coef cient 1 00000 001 Conversion coef cient 0 705479 002 Conversion coef ci...

Page 12: ...01 Enabled 80 Internal resolution 000 Do not use 001 255 81 1 timer ms 000 Do not use 001 255 82 3 timer ms 000 Do not use 001 255 91 Mechanical zero 1 000 255 Automatically set at span adjustment 92...

Page 13: ...ent 003 002 003 003 65 Zero point range 019 019 019 019 66 Zero point range on plus side 012 012 012 012 67 Over scale 003 003 003 003 68 Display hold function 000 000 000 000 69 Hold release against...

Page 14: ...t 001 001 002 001 65 Zero point range 019 019 019 019 66 Zero point range on plus side 012 012 012 012 67 Over scale 003 003 003 003 68 Display hold function 000 000 000 000 69 Hold release against in...

Page 15: ...t 001 001 002 001 65 Zero point range 019 019 019 019 66 Zero point range on plus side 012 012 012 012 67 Over scale 003 003 003 003 68 Display hold function 000 000 000 000 69 Hold release against in...

Page 16: ...YCO 1049 02 03 PPC 200W Technical Manual V Wiring Diagram 14 Note The PPC 200W is not washdown with this option Note The PPC 200W is not washdown with this option...

Page 17: ...YCO 1049 02 03 PPC 200W Technical Manual VI Parts List 15...

Page 18: ...1 27 YAM 1249 200027 CPU Board Assembly 1 28 YAM 1249 200028 Platform Cover S S 1 29 YAM 1249 200029 Level Indicator 1 30 YAM 1249 200030 Sealing Screw 1 32 YAM 1249 200032 Upper Housing Assembly Inc...

Reviews: