YCO-1049-02-03
PPC-200W
Technical Manual
II.B Test Mode
II Set-up Procedures
1. Re move the scale plat form.
2. Re move the fi ve screws and one lead wire seal screw from the
un der car riage of the scale.
3. Gently lift and re move the scale cover and re place the scale plat-
form.
4. Turn the scale on and short the two “TEST” jump ers on the left of
the PCB above the LCD.
5. Once the scale is in Test Mode, the dis play can be ro tat ed through
the fol low ing func tions by press ing
:
Internal Count
Initial Count
Direct Raw Count
Average Raw Count
Battery Check, A/D Con ver sion
ROM Ver sion
Display Seg ment Check
6. Exit Test Mode by press ing
.
+
- decrease value by ten
4. Press
to save changes and press
to exit User
parameter.
I Set-Up Procedures
II.A User Mode (con’t)
2