YCO-1049-02-03
PPC-200W
Technical Manual
II Set-up Procedures
3
II.C System Parameter Mode
1. Enter Test Mode (see section II.A for details.)
2. Press
and
to enter System Parameter Mode. The
scale will indicate a keyword fol lowed by it’s value.
3. To select the desired keyword, use the keys as follows:
- increase keyword by one
+
- decrease keyword by one
Note: Use the System Parameter List in Section IV.B to locate
the function to be changed and it’s standard value.
Example: To select zero tracking time press
eight
times.
4. User parameters are also set in the System Parameter Mode.
Press
to cycle through system keywords 30 through 40.
After system keyword 40, user keyword 01 is displayed. Press ing
will continue to cycle through user keywords 01 through 09
and then return to system keyword 30.
Note: Use the User Parameter List in Section IV.A to locate the
function to be changed and it’s standard value.
5. To change a parameter value, use the keys as follows:
- increase value by one
+
- increase value by ten
- decrease value by one
+
- decrease value by ten
Key word
Parameter Value