YCO-1049-02-03
PPC-200W
Technical Manual
8. Press
to save the new calibration points and return
to Test Mode. Internal counts should read 20000 +/- 3.
9. Press
at any time in the span-adjustment to exit to
Test Mode without saving any new calibration points.
10. Press
then
to return to normal operating
mode.
II Set-up Procedures
II.D Calibration Mode
II.E Factory Parameter Mode
1. Enter Test Mode (see section II.A for details.)
2. Enter Calibration Mode (see section II.C for details.)
3. Press
and
to enter Factory Parameter Mode.
The steps are the same as in System Parameter Mode
(see Section II.B) with the exception that the keywords
progress from 51 through 99.
Note: Use the Factory Parameter List in Sec tion IV.C
to locate the function to be changed. Use the Stan dard
Factory Parameter Settings List in Section IV.D to de-
ter mine the appropriate parameter value to enter.