Installation
Power Requirements
Ground
To prevent shock from electric current leak, be sure to connect the ground lead.
Contact Yamato Scientific or an electrician if a power source with a ground terminal is needed.
Never connect ground to with gas pipe, water pipe, telephone line, or lighting conductor.
SM200/300 SM310/510
Ground outlet is recommended for use.
No power plug is attached to these types of units.
Ground properly according to power source.
Connect to sufficient power source
Type Power
Source
Necessary
Capacity
SM200
AC115V
single-phase
13A or more
SM300
AC115V
single-phase
17A or more
SM310
AC220V
single-phase
11A or more
Be sure to connect the power cord to
sufficient power source.
(See right table)
SM510
AC220V
single-phase
11A or more
Attaching the power cord
Be sure that the breaker is turned off prior to connecting the power cord. The power cord plug is
not attached to the SM310 and 510. Select an adequate capacity plug and terminal according
to the connecting power source.
Green (Ground Lead)
Black (Power Side)
White (Power Side)
Power plug
Ground outlet
Ground lead