4. COMMUNICATION REFERENCE TABLE
Reference
No.
FNC
Code
R/W
Data name
Setting range
(in-transmission)
Initial
value
Remarks
41014
03
06
R
W
Defrost operation cycle
setting
1-96h
(1-96)
24h
Automatic operation only*
41015
03
06
R
W
Defrost operation time
setting
0-59min
(0-59)
5min
Common to automatic &
manual operation*
41016
03
06
R
W
Fan speed setting:
during standby
0-10
(0-10)
10
*
41017
03
06
R
W
Fan speed setting:
during operation
0-10
(0-10)
10
*
41018
03
06
R
W
Fan speed setting:
at the end of operation
0-10
(0-10)
10
*
41019
03
06
R
W
Auto damper setting:
damper aperture during
standby
0/25/50/75/100
%
(0/1/2/3/4)
0%
*
41020
03
06
R
W
Auto damper setting:
damper aperture during
operation
0/25/50/75/100
%
(0/1/2/3/4)
0%
*
41021
03
06
R
W
Auto damper setting:
damper aperture at the
end of operation
0/25/50/75/100
%
(0/1/2/3/4)
0%
*
41022
03
06
R
W
Event output setting:
event 1 output during
standby
OFF/ON
(0/1)
OFF
*
41023
03
06
R
W
Event output setting:
event 1 output during
operation
OFF/ON
(0/1)
OFF
*
41024
03
06
R
W
Event output setting:
event 1 output at the
end of operation
OFF/ON
(0/1)
OFF
*
41025
03
06
R
W
Event output setting:
event 2 output during
standby
OFF/ON
(0/1)
OFF
*
41026
03
06
R
W
Event output setting:
event 2 output during
operation
OFF/ON
(0/1)
OFF
*