6
3
Press
[TARE]
to select the
digit bit;
Press
[ZERO]
to change
the value;
Press
[#]
for confirm the
input of full capacity
[FULL ]
Full capacity setting
For example: Press
[#]
when displayed
【
025000
】
,
then the full capacity is set to be “25000”, and the
indicator enters zero point calibration
4
Make sure there is no load
on scale, and press
[#]
when the stable indication
sign is on
[nOLOAD]
Zero point calibration
5
Add full capacity load, press
[TARE]
to continue
[AdLOAD]
Calibrate the full capacity
For example:
Load 25000 weight on scale(as we set in step 3)
Use
[TARE]
and
[ZERO]
to change the value to
be 25000.
When stable indication sign is on, press
[#]
to
confirm.
Press
[TARE]
to select the
digit bit;
Press
[ZERO]
to change
the value accordingly with
the full capacity;
Press
[#]
to confirm when
the stable indication sign is
on
[025000]
6
Press the calibration switch
at the back housing of the
indicator.
[ End]
It saves the calibration parameter and back to the
weighing mode.
Note
: if no pressing the calibration switch at back of
indicator, all the parameters won’t be saved.
3.2 FAST CALIBRATION
Connect load cell properly, then turn on the indicator, keep pressing
[ # ]
key while its
initialization, it will enter into the calibration mode, and display
【
d X
】
.
3.2.1 Fast calibration of zero point
At any time before it showing
[nOLOAD]
, press
[FUNC]
. Indicator will keep the original
division, decimal point, full capacity, and enter the zero point calibration directly. Making
sure there is no load on the scale, and the stable indication sign is on, press
[ZERO]
to
re-calibrate the zero point.
The indicator will display
[End]
. Press the calibration switch at
the back of the indicator to save the setting and get back to the weighing status.
3.2.2 Fast calibration of full capacity
At any time before it showing
[AdLOAD]
, press
[*]
. It keeps the original division, decimal
point, full capacity, zero point, and enters into the full capacity calibration directly. After the
full capacity is reset, press the calibration switch at the back of the indicator to save the
setting and get back to the weighing status.