7
Chapter 4 Operation
4.1 POWER ON AND AUTO ZERO-SETTING
4.1.1 The indicator will perform “000000-999999” to self-checking when turning on. Then it
will enter weighing mode.
4.1.2 When power on, if loading weight on the scale deviates from the zero point, but still
within zero set range, the indicator will set zero automatically; if out of range, it is
necessary to adjust the zero point or recalibrate or reset.
4.2 MANUAL ZERO SETTING (AUTOMATICALLY)
4.2.1 In weighing mode, when there is some error when unloaded, press [Zero] to make
the indicator to be zero.
4.2.2 If the displayed value deviates from zero point, but still within zero-range, pressing
[Zero] key is available. Otherwise, [Zero] key is invalid. (In this status, please
recalibrate or reset zero parameters)
4.2.3 Only when stable annunciator is on, zero operation can be available.
4.3 TARE FUNCTION
When Indicator at weighing status, and displaying positive weight stable, press [ Tare ]
key, indicator will deduct the displayed weight value as tare weight. Then indicator
displays net weight as “0 ”, and Tare sign annunciator is on.
4.4 COUNTING FUNCTION
In weighing mode, press [Func] to enter the counting state, it will display “count”, and
press [*], it will display “C00000”, then press [Tare] to move the digit corresponding with
the small triangle, press [ZERO] key the digit where the triangle is will increase one, input
sample number, press [*] to enter into counting mode, the corresponding indicating
triangle light will be on. Press [Func] to return back to weighing mode.
In counting mode, the display will show “count”, press [*] twice, it will enter into counting
mode directly, indicator will calculating the sample and display it. (in this process, if “ERR
4” shows, it means the sample is failed, indicator will show last time’s sample.)
4.5 ACCUMULATING FUNCTION
In weighing mode, press [ * ] key, indicator will accumulate current weight, press [*] again,
it will return back to weighing mode.
In zero mode, press [*], it will show current accumulating value, in the accumulating mode,
press [Func] to zero the weight.
4.6 USER’S FUNCTION SETTING