X K 3 1 9 0
-
C 602
11
[******]
[0_Zone]
Zero zone
1. The discharge is deemed over
when the indicator shows the gross
weight is below zero zone.
2. The indicator can print and sum
only when the gross weight is
above zero zone.
12
[ ***]
[t0 ]
Feed measuring delay (
0.0 ~ 25.5
seconds)
Avoid misjudgment of weight
caused by the weight impact at
startup
13
[ ***]
[t1 ]
Measurement delay of quick feed
over
(0.0 ~ 25.5 seconds)
When t1 = 0, quick feed and slow
quick are activated at same time.
14
[ ***]
[t2 ]
Measurement delay of slow feed over
(0.0 ~ 25.5 seconds)
15
[ ***]
[t3 ]
Gradual feed output time (
0.0 ~ 25.5
seconds)
16
[ ***]
[t4 ]
Interval time of gradual feed
(0.0 ~
25.5 seconds)
17
[ ***]
[t5 ]
Qualified output time (
0.0 ~ 25.5
seconds)
18
[ ***]
[t6 ]
Discharge over delay (
0.0 ~ 25.5
seconds)
19
[ ***]
[t7 ]
Re-feed delay (
0.0 ~ 25.5 seconds)
20
[
A
]
[Print?]
A
—
whether to print parameter
setting
(0: no printing: 1: printing)