X K 3 1 9 0
-
C 602
The instrument cannot
enter running state.
Working parameter
setting has problems.
Please check whether
parameters set are
reasonable according to
specifications.
Cannot work in
accordance with
expected steps.
There are problems of
parameter setting.
The required input
signals for running
procedure is missed
Ditto
Check whether “allow
feed”, “allow
discharge” signals are
effective.
Cannot print
Wrong printer settings
Check printer setting
parameters and serial
port (1) setting
parameters
Unrecognized
characters occur during
printing
Wrong printer settings
Poor contact of print
cable
Check whether printing
setting parameters are
conforming to the
printer
Check printer cable and
plug or replace
printing cable
Figure flickering
Weighing platform is
unstable
Weighing platform is
vibrating
Transducer screening
wire is not connected
well
Electromagnetic
inference at site is too
large
AD shifting speed is
too quick
Filtration strength is too
small
Improve loader
structure
Adopt measures to
reduce weighing
platform vibration
Connect earthing wire
and screening wire
well
Reduce/decrease
electromagnetic
inference at site
Reduce AD shifting
speed (
【
SET 1
】
parameter3
B
)
Increase filtration
strength
(
【
SET 1
】
parameter
3
C
)
Figure response is slow. AD shifting speed is
slow
Filtration strength is too
large
Increase AD shifting
speed (
【
SET 1
】
parameter 3
B
)
Reduce filtration
strength