16-1
IM 2558A-01EN
Specifications
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
App
Index
16.1 Output
Output
AC Voltage
Range
Output range
Guaranteed
Accuracy Range
1
Resolution
Maximum Output
2
Output Resistance
3
100 mV
0 mV to 144.00 mV
1% to 120%
10 µV
―
Approx. 10 Ω
1 V
0 V to 1.4400 V
1% to 120%
100 µV
0.5 A or greater
Approx. 6 mΩ
10 V
0 to 14.400 V
1% to 120%
1 mV
Approx. 3 A
Approx. 6 mΩ
100 V
0 V to 144.00 V
1% to 120%
10 mV
Approx. 0.3 A
Approx. 6 mΩ
300 V
0 V to 432.0 V
1% to 120%
100 mV
Approx. 0.1 A
Approx. 6 mΩ
1000 V
0 V to 1440.0 V
1% to 120%
100 mV
Approx. 6 mA
Approx. 6 mΩ
AC Current
Range
Output range
Guaranteed
Accuracy Range
1
Resolution
Maximum Output
2
100 mA
0 mA to 144.00 mA
1% to 120%
10 µA
Approx. 15 V
1 A
0 A to 1.4400 A
1% to 120%
100 µA
Approx. 15 V
10 A
0 A to 14.400 A
1% to 120%
1 mA
Approx. 3 V
50 A
0 A to 72.00 A
1% to 120%
10 mA
Approx. 0.6 V
1
1% to 144% when the output frequency is 50 Hz or 60 Hz.
2
With the voltage or current at 1% to 120% of the range.
3
When <758933> or <758917> is in use; excluding aging and the effects of measurement leads.
Chapter 16 Specifications