2-2
IM 2558A-01EN
2.2 Output
CAUTION
When voltage or current is being generated, if a load that would cause the range generated
indicated below to be exceeded is connected, the instrument will detect the abnormal load
and turn off the output.
French
ATTENTION
Lors de la génération de la tension ou du courant, si la charge appliquée risque de sortir de
la plage générée indiquée ci-dessous, l’instrument détecte une charge anormale et coupe la
sortie.
Voltage Range and Range Generated
The maximum output at each voltage range is shown in the figure below. The maximum output
is 120% of each voltage range rating. If you set the voltage range to 300 V or 1000 V, the HIGH
VOLTAGE indicator will illuminate. If you set the voltage to 150 V or higher, a high-voltage caution
beep will sound.
1200 V
360 V
120 V
12 V
Voltage
6 mA
0.1 A
0.3 A
3 A
Current
100 mV 0 mV to 120.00 mV —
1 V
0 V to 1.2000 V
0.5 A or higher
10 V
0 V to 12.000V
approx. 3 A
100V
0 V to 120.00 V
approx. 0.3 A
300 V
0 V to 360.0 V
approx. 0.1 A
1000 V 0 V to 1200.0 V
approx. 6 mA
Range generated
Voltage
Range
Range
Generated
1
Maximum
Output
2
1 Generates 144% of the range when used
with deviation (–20%)
2 At 1% to 120% of the range
Current Range and Range Generated
The maximum output at each current range is shown in the figure below. The maximum output is 120%
of each current range rating.
60 A
12 A
1.2 A
Current
0.6 V
3 V
15 V
Voltage
100 mA 0 mA to 120.00 mA approx. 15 V
1 A
0 A to 1.2000 A
approx. 15 V
10 A
0 A to 12.000 A
approx. 3 V
50 A
0 A to 120.00 A
approx. 0.6 V
Range generated
Current
Range
Range
Generated
1
Maximum
Output
2
1 Generates 144% of the range when used
with deviation (–20%)
2 At 1% to 120% of the range