2.7 Other Features
Turning the Beep Sound On and Off
Beeps are used to indicate the operation status of the device. You can turn it on and off from the
SETUP menu.
You can turn the beep sound on or off for the following notifications. You cannot change the volume.
• When an error is detected during a self-test executed at power-on or executed manually (error
codes: E.911 to E950)
• When the voltage output is set to 150 V or higher
• When a sweep operation is held
You can not turn off the beep sound for the following notifications.
• When the cooling fan stops (error code: E.901)
• When an abnormal temperature is detected (error code: E.902 to E.903)
• When an internal power supply error is detected (error code: E.904)
Error Log Display
The error log keeps a record of error codes that occur in communication and self-tests while the 2558A
is on. You can display the error log from the SETUP menu. The error log is cleared when the power is
turned off.
Initialization
You can initialize the 2558A settings to their factory defaults. You can execute initialization from the
SETUP menu.
GP-IB and Ethernet settings are not initialized.
For a list of factory default settings, see appendix 5.
Product Information Display
You can view the firmware version, serial number, and so on. You can view the product information
from the SETUP menu.
• Firmware version
• Logic program version
• Boot program version
• Serial number