16-6
IM 2558A-01EN
16.6 General Specifications
Item
Specifications
Warm-up time
Approx. 30 minutes (after acclimatizing the 2558A to the operating environment for at least an hour)
Operating environment
Temperature:
5°C to 40°C
Humidity:
20% RH to 80% RH (no condensation)
Storage environment
Temperature:
–15°C to 60°C
Humidity:
20% RH to 80% RH (no condensation)
Operating altitude
2000 m or less
Installation location
Indoors
Orientation
Horizontal
Rated power supply voltage 100 VAC to 120 VAC, 200 VAC to 240 VAC
Permitted supply voltage
range
90 VAC to 132 VAC, 180 VAC to 264 VAC
Rated power supply
frequency
50 Hz/60 Hz
Permitted power supply
frequency range
48 Hz to 63 Hz
Maximum power
consumption
Approx. 200 VA
Withstand voltage
1500 VAC for 1 minute between the power supply and case
External dimensions
132 H × 426 W × 400 D mm
Weight
Approx. 20 kg
Safety standards
Compliant standard
EN 61010-1, EN61010-2-030
Overvoltage Category II
1
Pollution degree 2
2
Measurement category; Other (O)
3
Emissions
Compliant standard
EN61326-1 Class A
EN55011 Class A, Group 1
EN61000-3-2
EN61000-3-3
RCM EN55011 Class A, Group 1
Korea Electromagnetic Conformity Standard (
한국 전자파적합성기준
)
This product is a Class A (for industrial environment) product. Operation of this product in a
residential area may cause radio interference in which case the user will be required to correct
the interference.
Cable conditions
Signal output terminals (OSC OUTPUT)
Use a BNC cable that is 3 m or less in length.
External signal input terminal (EXT OSC INPUT)
Use a BNC cable that is 3 m or less in length.
Voltage terminals (VOLTAGE)
Use the included measurement leads (B8506ZK; red and black, one pc. each; length: 1 m).
Current terminal (CURRENT)
Use the included measurement leads (B8506WA; red and black, one pc. each; length: 1.5 m).
GP-IB connector (GP-IB)
Use a shielded cable that is 3 m in length or less.
Ethernet port (ETHERNET)
Use a category 5 or better Ethernet cable (STP) that is 30 m in length or less.
USB port for PCs (USB)
Use a shielded USB cable that is 3 m in length or less.
Immunity
Compliant standard
EN61326-1 Table 2 (for industrial locations)
Influence in the immunity test environment : Within ±15% of the range setting
Cable conditions
Same as the cable conditions for emission above.
Environmental standard
Compliant standard
EN50581 Monitoring and control instruments
1 The overvoltage category (installation category) is a value used to define the transient overvoltage condition and includes the
rated impulse withstand voltage. The overvoltage category II applies to electrical equipment that is powered through a fixed
installation, such as a switchboard.