IM CL235

 

9

 

2.10 Average Measurement 

(1)  Set the Function Selector switch to the "

A", "

A", "

V" or "

V" 

position. 

(2) Press the 

  button once to enter from the normal mode to the Aver-

age mode. "

" should be indicated on the display. 

(3)  Follow instructions for ACV, DCV, ACA or DCA measurement. 
(4)  The display shows a running average of six readings over an interval of 

about 2 seconds. 

 

 

10 

IM CL235

3. Other Functions 

3.1 Sleep 

Function 

This is a function to prevent the instrument from being left powered on in 
order to conserve battery life. This function causes the instrument to enter the 
Sleep (powered-down) mode about 30 minutes after the last switch or button 
operation. 
To exit the Sleep mode, turn the Function Selector switch back to "OFF", then 
to any other operation, or press any button. 
The current is consumed a little in the Sleep mode.

 

 

3.2  Data Hold Function 

This is a function used to freeze the measured value on the display. Press the 
Data Hold button to freeze the reading. The reading will be held regardless of 
subsequent variation in input. "

" is shown on the upper right corner of the 

display while the instrument is in the Data Hold mode. 
To exit the Data Hold mode, press the Data Hold button again. 

NOTE 

If the instrument in the Data Hold mode goes into "sleep", it will return to the 
normal mode. 

 

3.3 LoHz 

Function 

In ACV or ACA range, if frequency of the voltage or current under test is 40Hz 
or lower, the display indicates "LoHz" and sample rate is automatically 
switched from the normal 3 times/sec to 2 times/sec to reduce fluctuation of 
the reading. 
"LoHz" is also indicated where input is 3% of full scale or less. 

 

3.4 Optional 

Accessories 

Clamp Adapter Model 99025 (For AC current measurement only) 
Clamp Adapter Model 99025 is designed to increase the measuring capability 
of a clamp meter. With the use of the Clamp Adapter, you can not only extend 
current range over 3000A, but also clamp on a large bus-bar or conductor. 
(1)  Set the Function Selector switch to the "

A " position. 

(2)  As shown in the figure below, clamp Model CL235 onto the pickup coil of 

Model 99025. 

(3)  Clamp Model 99025 onto the bus-bar or conductor under test. 
(4)  Take the reading on Model CL235 and multiply it by 10. 

 

IM CL235

 11

 

 

NOTE 

For the detailed specification, refer to the Clamp Adapter User’s Manual. 

 

 

12 

IM CL235

4. Battery Replacement 

WARNING

To avoid electric shock hazard, make sure to set the Function Selector switch 
to "OFF" and remove the test leads from the instrument before trying to re-
place battery. 

 

CAUTIOIN

Make sure to install battery in correct polarity as indicated in battery com-
partment. 

 

If the battery voltage becomes too low for the instrument to operate normally, 
"

" is shown on the display. Then, replace the battery. Note that when 

the battery is completely exhausted, the display blanks without "

shown. 
(1)  Set the Function Selector switch to the "OFF" position. 
(2)  Unscrew and remove the battery compartment on the bottom of the in-

strument. 

(3)  Replace the battery observing correct polarity. Use a new 6F22 (Alkaline) 

or 6LR61 (Manganese) 9V battery. 

(4)  Re-place and screw the battery compartment cover. 

Battery

compartment cover

Screw

Battery

 

 

Summary of Contents for CL235

Page 1: ... Indicates that there is a possibility of serious personal injury of damage to the instrument if the operating procedure is not followed correctly and describes the precautions for avoiding such injury or damage NOTE Draws attention to information essential for understanding the op eration and features ii IM CL235 WARNING Never make measurement on a circuit above 600V AC or 600V DC Do not use the ...

Page 2: ...h to the OFF position after use When the instrument will not be in use for a long period of time Place it in storage after removing the battery Use a damp cloth and detergent for cleaning the instrument Do not use abrasives or solvents NOTE Radiation immunity affects the accuracy of CL235 testers under the condi tions specified in IEC801 3 1993 If equipment generating strong electromagnetic interf...

Page 3: ...t WARNING Do not make measurement on a circuit above 600VDC This may cause shock hazard or damage to the instrument or equipment under test Do not make measurement with the test leads plugged into the instrument Correct Wrong 1 Set the Function Selector switch to the A position DC should be shown on the upper left corner of the display 2 With the transformer jaws closed and without clamping them o...

Page 4: ... LCD display 3 With the tip of the test leads shorted together press the button to offset the resistance of the test leads 4 Press the button once to enter from the normal mode to the conti nuity check mode should be indicated on the display 5 Connect the tip of the test leads to the circuit under test If the resistance is 20Ω or less the buzzer beeps 2 8 Frequency Measurement On ACA or ACV range ...

Page 5: ...witched from the normal 3 times sec to 2 times sec to reduce fluctuation of the reading LoHz is also indicated where input is 3 of full scale or less 3 4 Optional Accessories Clamp Adapter Model 99025 For AC current measurement only Clamp Adapter Model 99025 is designed to increase the measuring capability of a clamp meter With the use of the Clamp Adapter you can not only extend current range ove...

Page 6: ... The scales are however actually calibrated in terms of the effective value of a sine wave though the clamp meter is responding to the average value The calibration is done with a conversion factor of 1 111 for sine wave which is found by dividing the effective value by the average value These instruments are therefore in error if the input voltage or current has some other shape than sine wave Wa...

Page 7: ...r still fails to operate correctly and needs repair con tact the vendor from whom you purchased the instrument or the nearest Yokogawa M C sales office Turn off the POWER switch once then turn it back on again If the tester does not turn on replace the battery with a new one Calibration It is recommended that the instrument be calibrated once every year ...

Page 8: ...IA PTY LTD AUSTRALIA Phone 61 2 9805 0699 Facsimile 61 2 9888 1844 YOKOGAWA BLUE STAR LTD INDIA Phone 91 80 227 1513 Facsimile 91 80 227 4270 LTD YOKOGAWA ELECTRIC RUSSIAN FEDERATION Phone 7 095 737 7868 Facsimile 7 095 737 7869 計 測 機 器 営 業 部 180 8750 東京都武蔵野市中町 2 9 32 電話 0422 52 5984 ファクシミリ 0422 55 8953 関 西 支 店 564 0063 大阪府吹田市江坂町 1 23 101 大同生命江坂ビル 10 階 電話 06 6368 7041 ファクシミリ 06 6368 7045 中 部 支 店 4...

Reviews: