7
IM 701430-51E
Turn ON/OFF the HF Rejection (HF Rej)
Specify 15 kHz or 20 MHz if you wish to use a signal that is obtained by removing the
high frequency components (frequency components greater than 15 kHz or 20 MHz)
from the trigger source signal as the trigger source.
Setting the Hold Off
See section 6.4, “Setting the Hold Off Time” in the
DL7100/DL7200 User’s Manual
(IM701410-01E).
Operating Procedure
Setting the Trigger Type
1.
Press the “ENHANCED” key.
2.
Press the “Type” soft key displays the trigger type selection menu.
3.
Press the “CAN Bus” soft key.
Setting the Trigger Pattern
4.
Press the “Set Pattern” soft key to display a dialog box for setting the trigger
pattern.
5.
Turn the jog shuttle to move the cursor to “Bit Rate.”
6.
Press the “SELECT” key.
7.
Turn the jog shuttle to select “1M,” “500k,” “250k,” “125k,” “100k,” “83.3k,” “50k,”
or “33.3k.” Press the “SELECT” key to confirm the setting. Pressing the
“RESET” key will reset the value to 500k.
8.
Turn the jog shuttle to move the cursor to “Sample Point.”
9.
Press the “SELECT” key.
10.
Turn the jog shuttle to set the sample point. Press the “SELECT” key to confirm
the setting. Pressing the “RESET” key will reset the value to 62.5.
2. Setting the Trigger Conditions