16
IM 701430-51E
Displaying the Search Result
• When Indefinite Data Is Searched
The head of the indefinite data is displayed at the Zoom Position.
• When Patterns Are Searched
• When multiple pattern items (Identifier, RTR, Data Field, Ack, and Error) are
selected:
The head of the specified field is displayed at the Zoom Position. If you specify multiple
patterns, the Zoom Position moves to the beginning of the field that was found last in the
time sequence. However, for Identifier and Data field, if the pattern is set to all Xs, it is
equivalent to not specifying a pattern. Thus, search is not performed.
• When all pattern items (Identifier, RTR, Data Field, Ack, and Error) are OFF:
A message “Pattern is not specified” is displayed.
Jumping to a Specified Field (Field Jump)
Moves the Zoom Box to a particular field within the current frame. The applicable fields
are the following five types.
Identifier
Control Field
Data Field
CRC
ACK
The Zoom Box moves to the head of the specified field.
Note
When the memory length is 16 Mwords on a DL7200 16M model, normal operation may not be
possible for frames near the ends of the main screen due to the limitation on the Zoom Position. The
Zoom Position may stick to the selectable boundary value when search or Field Jump is executed.
Operating Procedure
Showing the CAN Bus Signal Analysis Function Screen
1.
Press the “SHIFT” key to set the keys in the shifted condition.
Functions marked in purple on the panel become active.
2.
Press the “ZOOM” key.
3.
Press the “Type” soft key to display the search method selection menu.
4.
Press the soft key corresponding to “CAN Bus” to select the search method.
Note
•
The item to be analyzed is the history data of CH1.
•
When the search type is set to CAN Bus, the display format of the zoomed waveform (see
section 8.8 in IM701410-01E) may change. “Main,” “Z2 Only,” and “Main & Z2” change to
“Main & Z1.”
4. Analyzing and Searching