3-2
IM DM7560-02EN
3.2 Calibration
For this instrument to make accurate measurement, the regular calibration (after receiving it, charged
calibration and adjustment are executed at our site) is recommended.
For the regular calibration for the entire instrument, contact your nearest YOKOGAWA dealer. The
regular calibration is recommended once a year.
Note that the life of the battery for data backup is 5 years in the normal temperature. The battery
cannot be replaced by the customer.
In addition, the customer can make CALIBRATION (calibration and adjustment) for this instrument with
SYSTEM/TOOLS/CALIBRATION menu.
The performance of each function has the standard range and there may be deviation from the range
because of temporal change. CALIBRATION is done in such a case.
Section 5.3 in the IM DM7560-01EN describes the calibration (adjustment) by CALIBRATION menu.