4-4
IM DM7560-02EN
4.1 Common Specifications
Signal
timing
Approx. 1.4ms
4ms or more
Judgment result
COMPLETE
GO/HI/LO
Warm-up time
1 hour after power up
Installation
Only indoor use
Operation environment
Ambient temperature/humidity
0°C to 50°C
(40°C and no dew allowed below the moisture amount of
80%RH.)
Altitude
2000 m or less(25°C or less)
Storage temp./humid.
-20°C to +60°C
(40°C and no dew allowed below the moisture amount of
90%RH.)
Power supply
DM7560-1: AC100 ±10%, 50 Hz /60 Hz
DM7560-3: AC115 ±10%, 50 Hz /60 Hz
DM7560-6: AC220 ±10%, 50 Hz /60 Hz
DM7560-8: AC240 ±10%, 50 Hz /60 Hz
Power consumption
21 VA or less (option included)
Withstand voltage
DC±500 V (between LO terminal and ground earth)
Installation (over voltage)
Category II
category
(local level, electric product and portable product)
Dimension
225(W)×100(H)×366(D) mm (protuberance such as leg,
handle and knob excluded)
Weight
Approx. 3.0kg (protector and option included)
Expected life
LCD 50%
reduction of LED back light brightness around 70,000 hours
Relay
Approx. 10 million times (at normal use condition without
overload)
Approx. 100, 000 times (at max. overload of 1,000 V applied
voltage)
Data backup battery
5 years
Note) These attach to the articles of consumption, and the
exchange becomes a repair for a fee treatment.
Safety standard
Compliant standards
EN61010-1, EN61010-2-030
• Overvoltage category (installation category) II
1
• Measurement Category: II
2
(300 V), O(1100 Vpeak)
• Pollution degree 2
3
1 The overvoltage category (installation category) is a value used to define the transient overvoltage
condition and includes the rated impulse withstand voltage. Category I applies to electric equipment
whose power is supplied from a circuit that incorporates withstand voltage control. Category II applies
to electrical equipment that is powered through a fixed installation, such as a switchboard.
2 The measurement category of this instrument’s signal input terminals varies depending on the modules
that are installed. Use the instrument within the scope of the measurement category that corresponds
to the module specifications. Do not use the instrument outside the scope of the measurement
category that corresponds to the module specifications. The scope of each measurement category is
as follows.