4-15
IM DM7560-02EN
Specifications
4
4.5
Log function
Log mode
Switchable between 2 modes; i.e. NORMAL and BULK
Data size
NORMAL mode
100 k Readings fixed
BULK mode
Selectable from;
1 k, 2 k, 5 k, 10 k, 20 k, 50 k, and 100 k (Units: Readings)
Data to be Saved
The following are saved:
• Measurement data
• Logged date and time
• Name of each function
• Configuration information of each function
* The calculation name is displayed when NULL, dB, or SCALING
calculation is set on.
Export function
Outputting the measurement data stored in the LOG memory to the
USB memory.
File format
Text file
Saved data
Measurement data of function
Logging time
ON/ OFF can be set.
* When set ON, the date and time are saved.
Format
YYYY/MM/DD HH:mm:SS, xxxxxx
* x: unit of μsec
Attribute information
ON/ OFF can be set.
* The calculation name in the ON status (i.e. NULL, dB, or SCALING)
is saved.
4.5.1
NORMAL mode
It is a mode stored in the LOG memory while in real time monitoring the measurement data.
The sampling rate is not guaranteed.
After acquiring log data count
When the number of acquired data pieces exceeds 100 k
Readings, the mode continues by the FIFO* operation.
* FIFO: First In, First Out (First-in data is processed in
sequence.)
Log data clear
It is cleared at the following.
• INITIALIZE action by DISPLAY menu (M5 key is pressed)
• PUSH action when the rotary knob is invalid
• Additionally, refer to the clear condition table of each data of
section 6.4 in the use's manual(IM DM7560-01EN).