<3. Installation>
35
IM 01C25A01-01E
3.2.1 EJ 210
F0305.ai
Gasket
Stud bolt
Nut
Figure 3.5
EJ 210 Mounting
3.2.2 EJ 118 and EJ 438
Nut
Flange
Diaphragm
ød
Gasket
F0306.ai
Bolt
The product is
shipped with these
parts assembled.
Correctly install the diaphragm
seals on the high and low pressure
sides of the process (The label on
each diaphragm seal is marked
HIGH or LOW).
Figure 3.6
Mounting the Diaphragm Seals
3.3 Diaphragm Seals Installation
Consideration
IMPORTANT
F0307.ai
50 mm minimum
Minimum
liquid level
High
pressure
side
Low
pressure
side
The transmitter should
be installed as low as
possible below the
position where the high
pressure side
diaphragm seal is
installed.
Install the sealed diaphragm
so that the shank positions
downward.
IMPORTANT
Figure 3.7
Installing the Diaphragm Seals to a
Tank