<7. Operation>
7-1
IM 01C27C01-01EN
7. Operation
7.1 Preparation for Starting
Operation
The EJX210B flange mounted differential pressure
transmitter measures the levels or densities of
liquids. This section describes the operation
procedure for the EJX210B as shown in Figure 7.1
when measuring a liquid level in an open tank.
NOTE
It is required to set security and network
information to enable the transmitter to be
connected to the Field Wireless Network. For
more details, refer to section 7.4 “Connecting to
the Field Wireless Network”.
(a) Confirm that there is no leak in the connecting
part of the transmitter mounting flange.
Remove the plastic dust cap placed in the
process connector (low pressure side).
(b) Insert batteries into the battery case, and install
to the transmitter. To insert batteries into the
battery case, be careful to polarity of batteries
and battery case. For details of Installation of
battery, refer to section 9.4.6 and 9.4.7.
For the external powered type, turn on the
power and connect wireless field instruments to
field wireless network.
Battery case is installed in the transmitter when
shipped from the factory, however, batteries are
sold separately and not included.
(c) Using the field wireless configuration tool or
the device configuration tool, confirm that
the transmitter is operating properly. Check
parameter values or change the setpoints as
necessary.
Integral Indicator’s indication can be used
to confirm that the transmitter is operating
properly. For details on how to confirm, refer to
subsection 8.4 “Self-Diagnostics.”
ISA100 devices display self-diagnostic information
in an easy-to-understand manner using four
categories (Check function, Maintenance required,
Failure, and Off specification) according to NAMUR
NE107*.
* NAMUR NE107 [ Self-Monitoring and Diagnosis of
Field Devices ]
F0701.ai
Open Tank
Figure 7.1
Liquid Level Measurement
Confirm that transmitter is operating properly
by integral indicator.
• If the transmitter is faulty, an error code is
displayed.
F0702.ai
Self-diagnostic error on integral indicator
(Faulity transmitter)
Figure 7.2
Integral Indicator with Error Code
NOTE
If any of the above errors are indicated on the
display of the integral indicator or the device
configuration tool, refer to subsection 9.5.3 for
corrective action.