10-26
IM DR231-01E
Selecting the temperature unit (TEMP)
The temperature unit can be selected from the following;
•
°
C which is the initial value, or
•
°
F
Note
• In the User’s manual all functions and operations are explained using the
°
C unit. When using the
°
F
unit, replace the
°
C unit with the
°
F unit. For the measurement range/measurement accuracy and
maximum resolution in case of using the
°
F range, refer to the following table as a replacement of the one
on page 14-2.
• Measurement range: Refer to the following table;
• Measurement accuracy/maximum resolution: Replace the values using the following formula
°
F = (
°
C value
×
9/5) + 32
• When changing the temperature unit as descibred above, other settings will be initialized at the same time
and therefore, all settings need to be done again. This initialization will be the same initialization as
described on page 9-17.
• In order to make the set or selected contents effective, the contents must be registered. For details, see
page 10-24.
• To return from the SET UP menu to the operation display mode, select END in the main menu. For
details, see page 10-24.
Measurement ranges for temperature measurements when using the unit
°
F
Input
TC
(Note that accuracy
of reference junction
compensation is not
considered.)
RTD
High resolution RTD
R
S
B
K
E
J
T
L
U
N
W
KPvsAu7Fe
Pt100(1mA)
Pt100(2mA)
JPt100(1mA)
JPt100(2mA)
Pt50(2mA)
Ni100(1mA)
SAMA
Ni100(1mA)DIN
Ni120(1mA)
J263*B
Cu10 GE
Cu10 L&N
Cu10 WEED
Cu10 BAILEY
Pt100(1mA)
Pt100(2mA)
JPt100(1mA)
JPt100(2mA)
32 to 3200°F
32 to 3200°F
32 to 3308°F
–328 to 2498°F
–328.0 to 1472.0°F
–328.0 to 2012.0°F
–328.0 to 752.0°F
–328.0 to 1652.0°F
–328.0 to 752.0°F
32 to 2372°F
32 to 4199°F
0.0 to 300.0K
–328.0 to 1112.0°F
–328.0 to 482.0°F
–328.0 to 1022.0°F
–328.0 to 482.0°F
–328.0 to 1022.0°F
–328.0 to 482.0°F
–76.0 to 356.0°F
–94.0 to 392.0°F
0.0 to 300.0K
–328.0 to 572.0°F
–220.0 to 302.0°F
–94.0 to 158.0°F
–220.0 to 302.0°F
–94.0 to 158.0°F
Type
Measurement Range
*
1
*
1
*
1
*
1
*
1
*
1
*
1
*
2
*
2
*
3
*
4
*
5
*
5
*
5
*
5
*
5
*
6
*
6
*
7
*
8
*
8
*
8
*
8
*
5
*
5
*
5
*
5
*1 :R, S, B, K, E, J, T :ANSI, IEC 584, DIN IEC 584, JIS C 1602-1981
*2 :L :Fe-CuNi, DIN43710, U :Cu-CuNi, DIN 43710
*3 :N :Nicrosil-Nisil, IEC 584, DIN IEC 584
*4 :W :W·5%RE-W·26%Re (Hoskins Mfg Co)
*5 :Pt50 :JIS C 1604-1981, JIS C 1606-1986
Pt100 :JIS C 1604-1989, JIS C 1606-1989, IEC 751, DIN IEC 751
JPt100 :JIS C 1604-1981, JIS C 1606-1989
*6 :SAMA/DIN
*7 :McGRAW EDISON COMPANY
*8 :Ranges to which accuracy applies :
Cu10 GE :–119.9 to 338.0°F, Cu10 L&N :–103.0 to 302.0°F,
Cu10 WEED :–4.0 to 482.0°F, Cu10 BAILEY :–4.0 to 482.0°F
10.11 Selecting the temperature unit from
°
C or
°
F (option)
Summary of Contents for DR231
Page 12: ...10 IM DR231 01E...