10-33
IM DR231-01E
Basic Settings (SET-UP)
10
If Power Failure Occurs While Report Function Is Active
The DR recorder takes different actions depending on the length of a power failure.
If the power failure time is longer than 12 hours:
The DR recorder makes a report immediately after it recovers from the power failure and then
stops making reports. It does not execute printing based on the settings for automatic printing.
Print out reports either using the communication function or from the FUNC menu.
Results of computing: The DR recorder computes data measured up to the point immediately
before the power failure.
Time of report making: The time when the power failure occurred.
If the power failure is less than 12 hours:
The DR recorder takes different actions depending on the time it recovers from the power
failure.
Time of Recovery from
After the Time of
Before the Time of
Power Failure
Report Making
Report Making
Condition after recovery from
Valid report function
Valid report function
power failure
(start of report making enabled)
(start of report making enabled)
Report making
Immediately after recovery from
Time for report making
power failure
Report printing
Immediately after recovery from
Time for report making
power failure
Data included in report making
Data measured up to the point of
Data measured over the given
power failure
period except the power failure time
If automatic report printing is turned on:
If the power failure time covers the given period of report making, no report for that period is made.
Handling of Faulty Data
If data being computed contain any faulty data, the DR recorder treats the data as summarized in
the following table, depending on the type of computing and faulty data.
Type of Faulty Data Average
Minimum/Maximum Instantaneous Value Sum
Positive overflow
Excluded from
Included in computing
Takes faulty data as
Excluded from
computing
the result of computing
computing
Negative overflow
Excluded from
Included in computing
Takes faulty data as
Excluded from
computing
the result of computing
computing
Channels included in
Excluded from
Excluded from computing Takes faulty data as
Excluded from
measurement set to SKIP
computing
the result of computing
computing
No channel included
Excluded from
Excluded from computing Takes faulty data as
Excluded from
in measurement
computing
the result of computing
computing
Error
Excluded from
Excluded from computing Takes faulty data as
Excluded from
computing
the result of computing
computing
Output of data disabled
Excluded from
Excluded from computing Takes faulty data as
Excluded from
computing
the result of computing
computing
The DR recorder prints the results of computing as shown in the following table if they are special
data.
Type of Data
Printout Format
Positive overflow
+******
Negative overflow
–******
Channels included in measurement set to SKIP
Space
No channel included in measurement
Space
Error
××××××
Output of data disabled
If data being computed contain faulty data, the DR recorder prints each digital data item,
beginning with an * or
×
as the status indication.
1. Power failure:
×
In the case of a power failure, the DR recorder prints the time for report making, also beginning
with an
×
.
2. Channels included in measurement set to SKIP/no channel included in measurement/error/
output of disabled data:
×
10.12 Working with the Report Function
Summary of Contents for DR231
Page 12: ...10 IM DR231 01E...