<2. Screens>
2-8
IM 12A01F01-03EN 2nd Edition : Feb. 19, 2019-00
2.4
Alarms
Present alarms in the converter/sensors are diagnosed and displayed on Home / Main screen.
NOTE
Backlight keeps flashing until all fails are cancelled.
While alarm is present, backlight does not turn off automatically.
Home Screen
Main Screen
B
B
B
A
A
A
Errors sample
A: Main alarm
, , ,
or
,
most critical alarm at the moment they happen, including sensors’ alarm
B: Sensor alarm
, ,
, or
When “Alarm NE107”* is “On”, (“On” as factory default), the icons in accordance with NAMUR
NE107*, one of
, , ,
lights on upon detection of each alarm.
If a sensor alarm was detected, one of
, ,
, lights on.
When “Alarm NE107” is “Off”,
only
(Fault) or
(Warning) blinks.
Icons shows details of the error status.
Icons
Fault
:
,
Touch the icon to see itemized error information. If you press the itemized details,
you will see cause and remedy. When "Burn" is set "Burn Low" or "Burn High", Fault
automatically performs "Burn."
Warning:
, , , , , ,
The symbol indicates condition that maintenance is required. Touch the icon to know
each error information. If you touch the itemized error status, you will see cause and
remedy.
For information on the alarm icons, see 4.7 Error settings.
Error descriptions are listed on FLXA 402 and each connected sensor
Home/Main screen > a Main alarm key on top (most serious alarm) > Converter alarm Detail
Home/Main screen > a Sensor alarm at bottom > Sensor alarm Detail