<5. Others (Converter)>
5-7
IM 12A01F01-03EN 2nd Edition : Feb. 19, 2019-00
5.1.2
Load configuration
You will load configuration data of all sensors connected to FLXA402 as current configuration
data. The screen of Load configuration shows present configuration set list. The figure below
shows an example.
Figure 5.8
Sample of Load configuration
When you tap one of the item (above), a dialog box appears.
Dialog box shows the configuration set you select. Here you cannot modify a configuration set file
name. Tap “Execute” to load the configuration file you select.
When the loading is completed successfully, FLXA402 returns to Home screen,
NOTE
Alarm dialog appears if FLXA402 does not identify the sensor, because different sensor is
connected for data loading than the one used for data saving.
Example
Configuration set information was saved in FLXA402 when both sensor connection of 1-1
and 1-2 were connected. If you load this configuration set while connecting only 1-2, an alarm
dialog pops up, because FLXA402 cannot load the configuration to 1-1, although it succeeds
loading onto 1-2.
In this case, “Load individual configuration file” (5.1.1) would work for loading onto the sensor
which was not connected while FLXA402 was saving the data.