<4. Setting converter>
4-22
IM 12A01F01-03EN 2nd Edition : Feb. 19, 2019-00
l
Favorite calibration
Available for only pH measurement. After setting a Favorite calibration, tap Favorite calibration
, then you can move directly to the calibration you configured here.
Option on Favorite calibration
pH Calibration
Manual zero (1 point)
pH Manual
zero (1 pt)
Manual zero/slope (2 point)
zero/slope(2 pts)
Manual zero/slope/ITP (3 pt)
zero/slope/ITP (3 pts)
Manual zero/slope 1,2 (3 point)
zerp/slope1,2 (3 pts)
Auto zero (1 point)
pH Automatic
zero (1 pt)
Auto zero/slope (2 point)
zero/slope(2 pts)
Auto zero/slope/ITP (3 point)
zero/slope/ITP (3 pts)
Auto zero/slope 1,2 (3 point)
zerp/slope1,2 (3 pts)
Off
Calibration *
*:
When Favorite calibration is Off, or for measurement other than pH, you will go to same screen as ”Calibration” leads you.
l
Conductivity Unit, Resistivity Unit,
For only SC or ISC measurement.
Conductivity Unit:
Auto, nS, µS, mS, S
Resistivity Unit:
Auto, Ω cm, kΩ cm, MΩ cm, GΩ cm
4.5.2
Script
You can display the arithmetic result based on the scripts, setting up to 5 characters to “Script unit
strings”, 0 to 3 digits to “Script number decimal places”.
NOTE
Latest revision does not support Script. Don’t configure Script.