<4. Setting converter>
4-34
IM 12A01F01-03EN 2nd Edition : Feb. 19, 2019-00
4.8
Logbook
Logbook of FLXA402 records history of events such as setting change, calibration, alarm on/off,
and others. Logbook helps you, for example, to be aware of maintenance due.
Data in Logbook cannot be deleted, once it is stored.
Configure Logbook setting, by selecting “On” or “Off” to each displayed item. (Default are all
“Off”.)
Once you store data, the data are sorted out, being categorized into Converter and sensors.
Alarm name
Description
Converter
Error on
Converter error occurs.
Error off
Converter error is cancelled
mA output settings
mA output settings are changed
Relay wash setup
Contact output Wash setup is changed.
Relay alarm setup
Contact output alarm setup is changed.
Relay function setup
Contact output function setup is changed.
Relay fail setup
Contact output error setup is changed.
Relay simulate setup
Contact output simulate setup is changed.
Sensor
Error on
Sensor error occurs.
Error off
Sensor error is cancelled.
Compensation setup
Compensation setup is chaged.
Calibration setup
Calibration setup is changed.
Impedance setup
Impedance setup is changed.
When you set “Error settings” to “Off”, the alarm does not come to display, but the converter
detects errors inside. You can see the error history recorded in the logbook, by setting “Logbook
setting” to “Error On”.
NOTE
SENCOM SA keeps the log to record alarm message, event history such as calibration.
The converter shows the event log only that took place after the corresponding sensor was
connected.
If you need to see logs, which was stored in SENCOM SA before the event took place, you
should use FieldMate.
NOTE
When the storage of Logbook is full, the old data gets deleted first