<6. CALORIE / PARAMETER SETTING>
6-23
IM 11T03E01-51E
CODE 13: Selection for Automatic Calibration
[*AUTO.C
]
This setting determines whether or not automatic calibration is carried out. If automatic calibration
is to be carried out, you must also configure CODE 15 at the service level.
Disable:
0
Enable:
1
[ENT]
*CODE
13
*SERVC
X
*AUTO.C
[ENT]
[YES]
F0615
*
1
*
2
*1:
See "Figure 6.4 Process at the Setting Level" on page 6-12.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
*
3
Figure 6.15
Selection for Automatic Calibration
CODE 14: Setting for Remote Semi-automatic Calibration
[*REMOT
]
This setting determines whether or not semi-automatic calibration is carried out remotely.
If remote semi-automatic calibration is to be carried out, you must wire the contact input.
You must also configure CODE 15 at the service level.
Disable:
0
Enable:
1
[ENT]
*CODE
14
*SERVC
X
*REMOT
[ENT]
[YES]
F0616.ai
*
1
*
2
*1:
See "Figure 6.4 Process at the Setting Level" on page 6-12.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
*
3
Figure 6.16
Setting for Remote Semi-automatic Calibration