<6. CALORIE / PARAMETER SETTING>
6-9
IM 11T03E01-51E
(2) Density/Calorie Conversion Factor Setting Mode
In order to this mode, select “MAN.CAL” in before page, and press [YES] key.
This mode is used to set a factor for converting density to calorie. This setting is mandatory when
measuring the calorie.
(a) Automatic zero-point adjustment [AT_ADJ] : 0 to 133.000, 0 to 3.5000
(b) Factor setting [SET_K] :
• Zero point [K0_ADJ] : -99999 to 99999
• K0 [K0] :
-99999 to 99999
• K1 [K1] :
0 to 99999
Note: The result of executing item (a) above is also reflected on [K0_ADJ] in item (b).
[ENT]
[ENT]
ADJUST
XXX.XXX
VALVE
END
X.XXXX
K0_ADJ
XXX.XXX
CAL.SET
X.XXXX
K0
X.XXXX
K1
Measurement mode
[YES]
[NO]
[NO]
[YES]
[ENT]
AT_ADJ
SET_K
[YES]
[YES]
[ENT]
[ENT]
F0602.ai
*2: Using the [ ], [ ] and [ENT] keys, set a value (numeral position of the decimal point and sign).
Each press of the [ ] key moves through the fields of the numeral decimal point and sign in this order.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
*3: Set a value using the [ ], [ ] and [ENT] keys.
*4 When the reading stabilizes, press the [ENT] key to confirm the factor.
This parameter does not accept input from the [ ] and [ ] keys.
[NO]
MAN.CAL
[YES]
ZERO
[NO]
SPAN
[NO]
*1: See "Figure 6.1 Modes of Operation Level" on page 6-8.
*
3
*
4
*
2
*
2
*
2
*
1
Figure 6.2
Density / Calorie Conversion Factor Setting Mode
Summary of Contents for GD402G /M1
Page 21: ...Blank Page ...
Page 37: ...Blank Page ...
Page 81: ...Blank Page ...
Page 115: ...Blank Page ...
Page 159: ...Blank Page ...
Page 167: ......
Page 169: ......
Page 171: ...Blank Page ...