7
IM 04L51B01-02EN
1
/C2 and /C3 cannot be specified together.
2
/D5 can be specified only for the GX20/GP20.
3
Only one option can be specified.
4
Only one option can be specified.
5
/UC40, /UC50, /US40, and /US50 cannot be specified for the GX10/GP10.
6
/CR20, /CR21, /CR40, and /CR41 cannot be specified for the GX10/GP10.
7
If /UC20 or /US20 is specified for the GX10/GP10, /CR11 cannot be specified.
8
If /UC30 or /US30 is specified for the GX10/GP10, /CR01, /CR10, and /CR11
cannot be specified.
9
A digital input module has M3 screw terminals.
10
The Display language is selectable from English, German, French, Russian,
Korean, Chinese, Japanese.
To confirm the current available languages, please visit the following website.
URL: www.yokogawa.com/ns/language/
11
Solid state relay type (Type Suffix Code: -U2).
12
Can be specified only for the GX20/GP20.
13
To connect an I/O base unit, you will need one I/O expansion module for the
GX/GP.
14
/MC option must be separately specified when the WT communication is
selected.
15
Optional code /MT (MATH) required if using the GX90XD’s or GX90WD’s pulse
input.
16
Selectable only when the power cord suffix code is D or F or R or Q or H or N.
17
Selectable only for the GP10 when the power cord suffix code is W.
18
The /MT option (computation) is required to perform pulse integration on
GX90XP pulse input modules.
19
If you want to write from a PLC to the GX/GP via EtherNet/IP communication, a
separate communication channel (/MC) is required.
20
If you want the GX/GP to load data from SLMP servers via SLMP
communication, a separate communication channel (/MC) is required.
21
If you want to load data from other devices into the GX/GP using Modbus client,
a communication channel (/MC) is required.
22
This is applicable only when a GX90UT PID Control Module is installed.
I/O Base Unit (Expandable I/O)
1
Model
Suffix Code
Description
GX60
I/O base unit
Type
-EX
I/O Expansion
Area
N
General
Power supply
1
100 VAC, 240 VAC
Power cord
D Power cord UL/CSA standard
F Power cord VDE standard
R Power cord AS standard
Q Power cord BS standard
H Power cord GB standard
N Power cord NBR standard
W Screw terminal, power cord not included
2
1
Include GX90EX (Expansion module), Stopper (antiskid rubber)
2
Intended use for panel or rack mounting only.
I/O Expansion Module (Expansion Module)
Model
Suffix Code
Description
GX90EX
I/O Expansion Module
Port
-02
2 ports
Type
-TP1
Twisted pair cable
-
N
Always N
Area
-N General
I/O Modules
GX90XA
Model
Suffix Code
Description
GX90XA
Analog Input Module
Channels
-04
4 channels (Type -H0 only)
-06
6 channels (Type -R1 only)
-10
10 channels (Type -C1, -L1, -U2, -T1 )
Type
-C1
Current, Scanner type
(isolated between channels)
-L1
DCV/TC/DI (400 VAC, 1 min),
Scanner type
(isolated between channels)
-U2
Universal, Solid state relay scanner
type (3-wire RTD b-terminal common)
-T1
DCV/TC/DI, Electromagnetic relay
scanner type (Isolated between
channels)
-H0
High-speed universal, individual A/D
type (isolated between channels)
-R1
4-wire RTD/resistance, scanner type
(isolated between channels)
-
N
Always N
Terminal type
-3
Screw terminal (M3)
-C
Clamp terminal
Area
N General
GX90XD
Model
Suffix Code
Description
GX90XD
Digital Input Module
1
Channels
-16
16 channels
Type
-11
Open collector/Non-voltage, contact
(shared common), Rated 5 VDC
-
N
Always N
Terminal type
-3
Screw terminal (M3)
-C
Clamp terminal
Area
N General
1
Optional code /MT (MATH) required if using the pulse input.
GX90YD
Model
Suffix Code
Description
GX90YD
Digital Output Module
Channels
-06
6 channels
Type
-11
Relay, SPDT(NO-C-NC)
-
N
Always N
Terminal type
-3
Screw terminal (M3)
Area
N General
GX90WD
Model
Suffix Code
Description
GX90WD
Digital Input/Output Module
1
Channels
-0806
Input 8 channels, Output 6 channels
Type
-01
Open collector/non-voltage contact
(shared common), rated 5 VDC;
Relay, SPDT (NO-C-NC)
-
N
Always N
Terminal type
-3
Screw terminal (M3)
Area
N General
1
Optional code /MT (MATH) required if using the pulse input.
Power Specialties, Inc. | www.powerspecialties.com | (816) 353-6550