6-5
IM 04L57B01-01JA
GX/GP/GM Communication Commands
1
2
3
4
5
6
App
p16
Input value of correction point 7
p17
Output value of correction point 7
p18
Input value of correction point 8
p19
Output value of correction point 8
p20
Input value of correction point 9
p21
Output value of correction point 9
p22
Input value of correction point 10
p23
Output value of correction point 10
p24
Input value of correction point 11
p25
Output value of correction point 11
p26
Input value of correction point 12
p27
Output value of correction point 12
Use Calibration Correction (Correction coefficient) (/
AH)
Syntax
SCalibCom,p1,p2,p3,p4,p5,p6,p7,p8,
p9,p10,p11,p12,p13,p14,p15,p16,p17
,p18,p19p,20p,p21,p22,p23,p24,p25,
p26,p27,p28,p29,p30,p31,p32,p33,p3
4,p35,p36,p37,p38,p39
p1
Channel number
p2
Mode
Correct
Correction Factor
p3
Number of correction points (2 to 12)
p4
Input value of uncorrected value 1
p5
Instrument correction coefficient 1
p6
Sensor correction coefficient 1
p7
Input value of uncorrected value 2
p8
Instrument correction coefficient 2
p9
Sensor correction coefficient 2
p10
Input value of uncorrected value 3
p11
Instrument correction coefficient 3
p12
Sensor correction coefficient 3
p13
Input value of uncorrected value 4
p14
Instrument correction coefficient 4
p15
Sensor correction coefficient 4
p16
Input value of uncorrected value 5
p17
Instrument correction coefficient 5
p18
Sensor correction coefficient 5
p19
Input value of uncorrected value 6
p20
Instrument correction coefficient 6
p21
Sensor correction coefficient 6
p22
Input value of uncorrected value 7
p23
Instrument correction coefficient 7
p24
Sensor correction coefficient 7
p25
Input value of uncorrected value 8
p26
Instrument correction coefficient 8
p27
Sensor correction coefficient 8
p28
Input value of uncorrected value 9
p29
Instrument correction coefficient 9
p30
Sensor correction coefficient 9
p31
Input value of uncorrected value 10
p32
Instrument correction coefficient 10
p33
Sensor correction coefficient 10
p34
Input value of uncorrected value 11
p35
Instrument correction coefficient 11
p36
Sensor correction coefficient 11
p37
Input value of uncorrected value 12
p38
Instrument correction coefficient 12
p39
Sensor correction coefficient 12
Query
SCalibCom[,p1]?
6.3 Setting Commands
SCalibUseCom
Calibration correction use on/off
Sets the calibration correction use of a communication
channel to on or off.
Syntax
SCalibUseCom,p1,p2
p1
Channel number
p2
Enable or disable (Off)
Off
Not Use
On
Use
Query
SCalibUseCom[,p1]?
Example
Enable the calibration correction on
communication channel 001.
SCalibUseCom,001,On
Description
• There is a limitation on the number of channels that p2
can be set to On.
Model
GX10
GP10
GX20-1
GP20-1
GM10-1
GX20-2
GP20-2
GM10-2
Number of channels
that can be set to On
50
150
300
• If p2=Off in the communication channel on/off setting
(SRangeCom), p2 is fixed to Off.
• For details on communication channels, see section
2.3.2, “Parameter Notation and Range” in the
Communication Command Manual.
SCalibCom
Calibration correction
Sets the calibration correction on communication
channels.
Disable Calibration Correction
Syntax
SCalibCom,p1,p2
p1
Channel number
p2
Linearizer mode (Off)
Use Calibration Correction (Linearizer approximation,
linearizer bias)
Syntax
SCalibCom,p1,p2,p3,p4,p5,p6,p7,p8,
p9,p10,p11,p12,p13,p14,p15,p16,p17
,p18,p19p,20p,p21,p22,p23,p24,p25,
p26,p27
p1
Channel number
p2
Mode
Appro
Linearizer approximation
Bias
Linearizer bias
p3
Number of correction points (2 to 12)
p4
Input value of correction point 1
p5
Output value of correction point 1
p6
Input value of correction point 2
p7
Output value of correction point 2
p8
Input value of correction point 3
p9
Output value of correction point 3
p10
Input value of correction point 4
p11
Output value of correction point 4
p12
Input value of correction point 5
p13
Output value of correction point 5
p14
Input value of correction point 6
p15
Output value of correction point 6
Summary of Contents for GX70SM
Page 1: ...User s Manual IM 04L57B01 01EN 1st Edition Model GX70SM Wireless Input Unit User s Manual ...
Page 2: ......
Page 12: ...Blank ...
Page 148: ...Blank ...
Page 168: ...Blank ...