1-4
IM 04L57B01-01EN
Range Details
Type
Range setting Range
Notes
TC
K
-200.0°C to 1370.0°C
-328 °F to 2498 °F
Type K
J
-200.0°C to 1100.0°C
-328.0 °F to 2012.0 °F
Type J
T
-200.0°C to 400.0°C
-328.0 °F to 752.0 °F
Type T
B
0.0°C to 1820.0°C
32 °F to 3308 °F
Type B
S
0.0°C to 1760.0°C
32 °F to 3200 °F
Type C
R
0.0°C to 1760.0°C
32 °F to 3200 °F
Type R
N
-270.0°C to 1300.0°C
-454 °F to 2372 °F
Type N
E
-200.0°C to 800.0°C
-328.0 °F to 1472.0 °F
Type E
WRe3-25
0.0 °C to 2400.0 °C
32 °F to 4352 °F
Type WRe (WRe3-25)
RTD
Pt100
-200.0°C to 600.0°C
-328.0 °F to 1112.0 °F
JPt100
-200.0°C to 550.0°C
-328.0 °F to 1022.0 °F
Voltage
20mV
-20.000 mV to 20.000 mV
60mV
-60.00 mV to 60.00 mV
200mV
-200.00 mV to 200.00 mV
2V
-2.0000 V to 2.0000 V
6V
-6.000 V to 6.000 V
10V
-10.000 V to 10.000 V
GS
0.4-2V
0.4000 V to 2.0000 V
1-5V
0.800 V to 5.200 V
DI
LVL
On (1)/Off (0) (voltage)
On: 2.5 V or more
Off: 2.3 V or less
DI
On (1)/Off (0) (contact)
Humidity
—
0.0 to 90.0% RH
1.2.3
Measurement Mode
There are two measurement modes.
Standard mode
Power frequency noise riding on the measured signal is rejected. Select the
power frequency for your region using the Wireless Input Unit Configurator.
The battery life (running time on battery) is shorter than in battery-save mode.
Battery-save mode
Power frequency noise riding on the measured signal is not rejected.
Depending on the measuring range, measurement errors will be large due to
the effects of noise, and the values may fluctuate.
The battery life (running time on battery) is 1.3 to 1.5 times longer than in
standard mode.
1.2 Functions
Summary of Contents for GX70SM
Page 1: ...User s Manual IM 04L57B01 01EN 1st Edition Model GX70SM Wireless Input Unit User s Manual ...
Page 2: ......
Page 12: ...Blank ...
Page 148: ...Blank ...
Page 168: ...Blank ...